ISO/DIS 4625-3
ISO/DIS 4625-3
ISO/DIS 4625-3: Binders for paints and varnishes — Determination of softening point — Part 3: Cup method (without a ball)

ISO/DIS 4625-3:2025(en)

ISO TC 35/WG 5

Secretariat: NEN

Date: 2025-05-02

Binders for paints and varnishes — Determination of softening point — Part 3: Cup method (without a ball)

© ISO 2025

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.

ISO copyright office

CP 401 • Ch. de Blandonnet 8

CH-1214 Vernier, Geneva

Phone: +41 22 749 01 11

Email: copyright@iso.org

Website: www.iso.org

Published in Switzerland

Contents

Foreword iv

Introduction v

1 Scope 1

2 Normative references 1

3 Terms and definitions 1

4 Principle 1

5 Apparatus 1

5.1 Softening point apparatus 1

5.2 Central processor 2

5.3 Measuring cell 2

5.4 Sample cup assembly 2

6 Calibration of apparatus using a primary standard 2

6.1 General 2

6.2 Reagent 2

6.3 Procedure 3

6.3.1 Filling the sample cup 3

6.3.2 Heating 3

6.3.3 Cleaning 3

6.3.4 Interpretation 3

7 Instrument check of apparatus using a secondary standard 3

8 Preparation of sample 4

8.1 Preparation 4

8.2 Filling the sample cup 4

9 Procedure 5

10 Expression of results 5

11 Precision and bias 5

11.1 General 5

11.2 Precision 5

11.3 Bias 5

12 Test report 6

Annex A (informative) Details of the interlaboratory comparison for the estimation of precision 7

A.1 General notes on the interlaboratory comparison 7

A.2 Samples 7

A.3 Determinations 7

A.4 Evaluation 8

Bibliography 1

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any patent rights identified during the development of the document will be in the Introduction and/or on the ISO list of patent declarations received (see www.iso.org/patents).

Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 35, Paints and varnishes.

A list of all parts in the ISO 4625 series can be found on the ISO website.

Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.

Introduction

The ISO 4625 standards series describes three widely used procedures for the measurement of the softening point of rosin-based resins, ISO 4625‑1,[1] the ring-and-ball method which was the accepted standard for many years and ISO 4625‑2,[2] the Mettler cup-and-ball method. This document adds a new method, the Mettler method without the ball.

ISO 4625‑1 is still the only standard test method accepted in regulatory documents such as 21CFR [3].

ISO 4625‑2 and this document are both called Mettler cup-and-ball methods. Although the recommended testing conditions differ the only difference between the equipment required is that this document does not use a ball. Surveys have shown that ISO 4625‑2 is the most widely used in the US and this document the most widely used in Europe. These methods are less time consuming than the ring-and-ball manual method and the equipment is less expensive than the ring-and-ball automated method.

As a consequence of the thermoplastic nature of the test resins, the softening points obtained using the recommended test conditions for all three methods are not the generally the same. Consequently, the test method and the testing conditions used should be noted in the final report.

Binders for paints and varnishes — Determination of softening point — Part 3: Cup method (without a ball)

1.0 Scope

This document specifies a method for determining the softening point of resins (including rosin) and can, under user-defined conditions, give results comparable to those obtained by ISO 4625-1 and ISO 4625-2.

2.0 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

ISO 4618, Paints and varnishes — Vocabulary

3.0 Terms and definitions

For the purposes of this document, the terms and definitions given in ISO 4618 and the following apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https://www.iso.org/obp

— IEC Electropedia: available at https://www.electropedia.org/

3.1

softening point

temperature at which the sample, suspended in a cylindrical cup with a 6,35 mm hole in the bottom and without a stainless steel ball, flows downward a distance of 19 mm, as the sample is heated at linear rate in air

4.0 Principle

In general, with materials of these types, softening does not take place at a definite temperature. As the temperature rises, these materials gradually change from brittle or exceedingly thick and slow-flowing materials to softer and less viscous liquids. For this reason, the determination of the softening point shall be made by a fixed, arbitrary, and closely defined method if the results obtained are to be comparable.

The disc of the test sample is placed in a cup and heated at a specified rate. The softening sample is forced by its own weight downwards. The temperature at which the sample has sunken by 19 mm is called "softening point".

5.0 Apparatus

5.1 Softening point apparatus

This commercially available instrument consists of control unit with a digital temperature indicator, matched measuring cell, cartridge assembly – consisting of sample cup and upper portion – and accessories. The control unit automatically regulates the heating rate of the measuring cell. The softening point is indicated on the readout, and the heating programme is stopped, when the sample flow triggers a detector.

5.1.1 Central processor

This unit shall provide a continuous, linear increase from 25 °C to 375 °C.

5.1.2 Measuring cell

This unit shall be capable of heating a sample cup assembly, as described in 5.4, at a linear rate with an accuracy of ±0,2 °C/min. It shall include a sensing system capable of detecting the softening point with a precision of 0,1 °C.

5.1.3 Sample cup assembly

A cup of chromium-plated brass and an upper portion, conforming to the dimensions shown in Figure 1.

Dimension in millimetres

Figure 1 — Cup

6.0 Calibration of apparatus using a primary standard

6.1 General

This step, required only occasionally, is designed to establish that the temperature indicated by the instrument is in agreement with a known standard. A special cup with a bottom orifice of 2,8 mm, is used instead of the one specified for testing the resin.

6.1.1 Reagent

Use analytical reagent of primary standard grade benzoic acid for the calibration. As this material is hygroscopic, it shall be stored in a tightly sealed container and replaced with fresh material from a newly opened supply if the melting point of the material is out of the expected range.

6.1.2 Procedure

6.1.3 Filling the sample cup

Place the cup on a clean, flat surface and fill it with benzoic acid crystals. Hand press the material into the cup with a 9,2 mm rod of dowel stock or equivalent. Refill and repeat the pressing step until the cup is filled with benzoic acid. Remove any crystals from the exterior of the cup.

6.1.4 Heating

Preheat the measuring cell to 121 °C, and maintain it at that temperature. Place the cartridge assembly containing the benzoic acid in position in the measuring cell. With the instrument set in the dropping point mode, start the required method, which includes a waiting period of 30 s to allow temperature equilibration between the measuring cell and sample followed by an automatic temperature ramp of 0,2 °C/min. The temperature will rise steadily at the correct rate until the melt point is reached, then remain steady on the readout.

6.1.5 Cleaning

Immediately remove the cartridge assembly. Check to determine that the sample has passed the detector and no pre-triggering has occurred. If a malfunction is suspected, the entire procedure shall be repeated. Inspect the apparatus carefully to ensure that no residue remains. Wash the cartridge parts with a suitable solvent to remove last traces of residue.

6.1.6 Interpretation

If the result is not (123,5 ± 0,5) °C, repeat the test. If the second value remains outside of the acceptable range, repeat the test using a fresh sample of benzoic acid. If the results remain outside of the acceptable range, the instrument requires recalibration or repair. Consult the manufacturer's instruction manual.

7.0 Instrument check of apparatus using a secondary standard

The primary calibration standard, benzoic acid, might not have a melting point similar to the softening point of the resin being tested. In such cases, it is desirable to calibrate the instrument using a secondary standard having a melting point similar to softening point of the test sample. Materials suitable for use as secondary standards are listed, but are not limited to those, in Table 1. Demonstration of conformance to the melting point of benzoic acid should be completed before a secondary standard is used.

Table 1 — Materials suitable as secondary standards

Approximate melting point

Material

°C

 

60

Stearyl alcohol

80

Diphenyl carbonate

100

Phenoxyacetic acid

4-t-Butylphenol

Methyl urea

110

m-Toluic acid

Resorcinol

120

Benzoic acid

130

Methyl 4-hydroxybenzoate

Urea

Benzamide

1,10-Decanedicarboxylic acid

140

Dimethyl terephthalate

o-Toluamide

Follow the same procedure as in 6.3.2 and 6.3.3 but start about 3 °C below the actual melting point of the secondary standard.

8.0 Preparation of sample

8.1 Preparation

Fold the tab on an oil-free disposable aluminium weighing dish to a horizontal position and pinch a pour spout into the dish directly opposite the tab. Weigh 5 g to 10 g of resin into the dish.

Place the dish and sample on a preheated hot plate, and cover with a 600 ml or larger beaker. Heat the sample until it is almost melted.

Remove the beaker from the hot plate, grasp the tab with forceps or needle-nose pliers, and stir the sample with a probe-type thermometer until it is completely melted and homogeneous. Avoid entraining air in the sample melt. The temperature of the sample should be no more than 60 °C above the expected softening point. Remove the thermometer and wipe clean. If the sample smokes or appears to boil, repeat the steps 8.1 to 8.3, setting the maximum temperature 10 °C less than the previous trial.

8.1.1 Filling the sample cup

Arrange two sample cups, small outlet down, on a piece of aluminium foil placed on a smooth, flat surface.

Grasping the dish tab with forceps or pliers, pour the molten resin into both of the sample cups. Fill the cups to the brim.

Allow the resin to cool for at least 10 min. Cleanly trim off the excess resin on the sample cup with a slightly heated knife or spatula, or grasp the cup in a pair of tongs and draw the top surface quickly and firmly over the surface of a heated metal plate. Do not make a determination on a sample which contains air bubbles.

Assemble the cartridge by placing the cup on the collector, then the upper portion on the cup.

Make sure that there is no sample on the outside of the cup, the inside of the jacket, or the collecting sleeve. These components shall be clean during analysis. The assembly should be wiped clean before inserting it into the measuring cell.

9.0 Procedure

The procedure described is for measuring the cup (without a ball) softening point of resins. The heating rate is usually determined by the user. A heating rate of 1,0 °C/min is typical for resins.

If a correlation is desired between cup (without a ball) and ring-and-ball softening point as described in ISO 4625-1 or cup-and-ball softening point as described in ISO 4625-2, a specific heating rate for that correlation shall be determined for each type of resin.

Preset the starting temperature of the central processor from 20 °C to 25 °C below the expected softening point of the sample and maintain this temperature for a few minutes. Place the cartridge assembly in the measuring cell. With the instrument in the softening point mode, start the determination, which includes a waiting period of 30 s to allow temperature equilibration between the measuring cell and sample followed by an automatic temperature ramp at the specified rate in °C/min. Heating will be maintained at this rate until the softening point is reached, after which a steady temperature will be displayed on the readout. Record this temperature as the softening point of the sample.

Remove the cartridge assembly upon completion of the test. Check to ensure that the sample has passed the detector and no pretriggering has occurred. If a malfunction is suspected, reject the result and repeat the test. Inspect the apparatus closely to ensure that no particles of residue remain.

Clean the sample cup by soaking in a suitable solvent.

Care shall be taken in cleaning the sample cup in order not to scratch or alter the size of the cup orifice.

Clean the interior of the cell immediately after removing the sample assembly. Use soft cotton swabs moistened in an appropriate solvent.

10.0 Expression of results

Calculate the mean of two determinations and round it to the nearest 0,5 °C.

11.0 Precision and bias

11.1 General

An interlaboratory comparison of the cup-and-ball softening point of seven resins with low and high softening points was run by 11 laboratories. Details of the round robin test are given in Annex A. The precision information given below is for the comparison of two test results.

11.1.1 Precision

Repeatability limit, 95 % (within laboratory):

0,2 °C

for low softening point

 

 

1,1 °C

for high softening point

Reproducibility limit, 95 % (between laboratories):

0,7 °C

for low softening point

 

 

2,3 °C

for high softening point

11.1.2 Bias

Since there is no accepted reference material, method or laboratory suitable for determining the bias for the procedure in this test method for measuring the cup (without a ball) softening point, no statement on bias is being made.

12.0 Test report

The test report shall include at least the following information

a) the type and identification of the product tested;

b) a reference to this document, i.e. ISO 4625‑3:—;

c) the heating rate used;

d) the result of the test, as indicated in Clause 10;

e) any deviation from the test procedure specified;

f) any unusual features (anomalies) observed during the test;

g) the date of the test.


  1. (informative)

    Details of the interlaboratory comparison for the estimation of precision
    1. General notes on the interlaboratory comparison

An interlaboratory comparison (ILC) has been carried out in 2014 and 2015 for the determination of precision using the cup-and-ball softening point.

11 laboratories participated in the comparison.

    1. Samples

For the interlaboratory comparison, seven resins with low and high softening points were tested – see Table A.1.

Table A.1 — Samples

Sample number

Sample

1

Gum rosin

2

Glycerol ester of rosin

3

Penta ester of rosin

4

Polyterpen resin

5

Setaprint 2404™

6

Escorez 1621™

7

Escorez 3405™

    1. Determinations

Five different determinations were carried out – see Table A.2.

Table A.2 — Determinations

Number

Code

Type of determination

1

RB

Ring and ball

2

Mettler softening point “cup and ball” method

2.1

SP1

                    with US steel balls, 2 °C/min

2.2

SP2

                    with metric steel ball, 2 °C/min

2.3

SP3

                    without steel ball, 1 °C/min

3

DSP

Mettler drop softening point without steel ball, 2 °C/min

    1. Evaluation

The calculation of the precision could be done based on a sufficient number of measurement results obtained. Detailed results and calculated values see Table A.3.

Table A.3 — Results and calculated values<Tbl_--></Tbl_-->

Determination

Gum rosin

Glycerol ester

Penta ester

Polyterpen resin

Setaprint 2404™

Escorez 1621™

Escorez 3405™

SP1

Number of laboratories

11

11

11

11

4

4

4

Mean value °C

83,62

83,14

96,38

107,94

119,10

128,56

137,88

Standard deviation, s

1,014 05

0,763 75

1,356 84

0,882 69

0,912 18

0,803 02

1,953 56

Repeatability, sr

0,374 04

0,179 57

0,269 73

0,272 98

0,650 96

0,324 17

0,590 48

Reproducibility, sR

1,059 16

0,800 22

1,722 51

0,923 40

0,995 85

0,911 37

2,234 20

SP2

Number of laboratories

11

11

11

11

4

4

4

Mean value °C

83,6

83,2

96,3

108,1

118,9

128,4

138,1

Standard deviation, s

1,174 28

0,825 26

1,197 18

0,834 88

1,549 32

0,660 76

1,949 85

Repeatability, sr

0,320 66

0,316 18

0,401 96

0,412 35

0,859 80

0,244 93

0,474 78

Reproducibility, sR

1,229 44

0,861 60

1,251 57

0,868 34

1,730 66

0,752 07

2,237 55

SP3

Number of laboratories

11

11

10

11

5

5

5

Mean value °C

83,7

83,2

96,8

108,0

118,6

128,3

138,0

Standard deviation, s

1,174 80

0,840 99

0,729 18

0,717 47

0,948 89

1,282 90

1,798 61

Repeatability, sr

0,280 80

0,245 34

0,227 67

0,389 57

0,607 32

1,147 74

0,593 09

Reproducibility, sR

1,230 82

0,880 14

0,765 49

0,744 73

1,039 99

1,347 95

2,031 79

RB

Number of laboratories

5

7

7

7

2

2

2

Mean value °C

81,9

83,8

96,5

106,9

117,8

126,4

135,9

Standard deviation, s

2,186 55

1,211 44

1,531 10

1,149 75

1,482 70

1,922 39

2,023 96

Repeatability, sr

0,415 29

0,419 98

0,485 99

0,570 95

0,878 45

0,618 04

0,398 96

Reproducibility, sR

2,437 27

1,299 11

1,643 83

1,223 53

1,976 05

2,713 84

2,897 39

Bibliography

[1] ISO 4625‑1, Binders for paints and varnishes — Determination of softening point — Part 1: Ring-and-ball method

[2] ISO 4625‑2, Binders for paints and varnishes — Determination of softening point — Part 2: Cup-and-ball method

[3] 21CFR – Code of Federal Regulations Title 21, Food and Drugs, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, MD 20993

espa-banner