ISO/DIS 18191:2025(en)
ISO/TC 147/SC 2
Secretariat: DIN
Date: 2025-04-22
Water quality — Determination of pHT in sea water — Method using the indicator dye m-cresol purple
Qualité de l'eau — Détermination du pHT dans l'eau de mer — Méthode utilisant l'indicateur coloré au pourpre de m-crésol
© ISO 2025
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
Contents Page
6.4 High‑quality spectrophotometer 4
6.5 Temperature-control system for spectrophotometer cell 4
6.6 System to warm samples to measurement temperature 4
9.1 Correction of measured absorbances 5
9.2 Calculation of the pHT of the sea water and indicator 6
Annex A (informative) Performance data 8
A.1 Interlaboratory comparison on TRIS buffer and natural sea water 8
A.2 Interlaboratory comparison on North Pacific sea water 10
Annex B (informative) Storage stability 12
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO [had/had not] received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 147, Water quality, Subcommittee SC 2, Physical, chemical and biochemical methods.
Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.
Introduction
The greenhouse effect induced by anthropogenic carbon dioxide, CO2, in the atmosphere is one of the serious global environmental issues. A key factor controlling the atmospheric CO2 is its absorption into the ocean. As a result of the absorption, the pH in the upper layer of the ocean is observed to have fallen gradually, and its influence on the living organisms is a matter of concern all over the world.
On the other hand, carbon capture and storage (CCS) technology is considered as a useful means of reducing the CO2 emissions from fossil fuel. When ocean environment such as sub-seabed aquifer is selected as a storage site, the monitoring of carbonate system including pH in sea water becomes very important. The analytical method for pHT in sea water (the total hydrogen ion concentration pH scale) samples requires specific conditions and techniques essential to the precise and accurate determination. This International Standard describes a method for the determination of pHT in sea water with the repeatability less than 0,003.
This method will provide international communities accurate data sets on pHT in sea water being compatible with each other. This is the base of national and international operational observation or monitoring programs of the oceanic carbonate system as well as individual research works.
Water quality — Determination of pHT in sea water — Method using the indicator dye m-cresol purple
WARNING — Persons using this document should be familiar with normal laboratory practice. This document does not purport to address all of the safety problems, if any, associated with its use. It is the responsibility of the user to establish appropriate safety and health practices and to ensure compliance with any national regulatory conditions.
IMPORTANT — It is absolutely essential that tests conducted in accordance with this document be carried out by suitably qualified staff.
1.0 Scope
Type text. This International Standard specifies a spectrophotometric determination of the pHT of sea water on the total hydrogen ion concentration pH scale. The total hydrogen ion concentration, [H+]t, is expressed as moles per kilogram of sea water. The method is suitable for assaying oceanic levels of pHT from 7,4 to 8,2 for normal sea water of practical salinity ranging from 20 to 40.
2.0 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 5667‑1, Water quality — Sampling — Part 1: Guidance on the design of sampling programmes and sampling techniques
ISO 10523, Water quality — Determination of pH
ISO 22719, Water quality — Determination of total alkalinity in sea water using high precision potentiometric titration
3.0 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https://www.iso.org/obp
— IEC Electropedia: available at https://www.electropedia.org/
3.1
total hydrogen ion concentration [H+]t
total hydrogen ion concentration of sea water, [H+]t includes the contribution of free hydrogen ions and the hydrogen ions bound to sulfate ions in the sea water and is defined as
(1)
where
[H+]F is the free concentration of hydrogen ion in sea water;
ST is the total sulfate concentration
KS is the acid dissociation constant for
The pHT is then defined as the negative of the base 10 logarithm of the hydrogen ion concentration as given in Equation (2):
(2)
3.2
practical salinity
S
practical salinity, symbol S, is a representative for the amount of dissolved salts in seawater and is defined as a polynomial function of the ratio R15 of the electrical conductivity of the sea water sample at the temperature of 15 °C on IPTS‑68 and the pressure of one standard atmosphere, to that of a potassium chloride (KCl) solution, in which the mass fraction of KCl is 32,4356 × 10-3, at the same temperature and pressure [xxx].
4.0 Principle
The values of pHT are determined by adding an indicator to sea water. For the sulfonephthalein indicators such as m‑cresol purple, the reaction of interest at sea water pHT is the second dissociation as given in Equation (3):
(3)
where I represents the indicator, which is present at a low level in a sea water sample. The total hydrogen ion concentration of the sample can then be determined as given in Equation (4):
(4)
The principle of this approach uses the fact that the different forms of the indicator have substantially different absorption spectra. Thus the information contained in the composite spectrum can be used to estimate [I2‑] / [HI-].
At an individual wavelength, λ, the measured absorbance, Aλ, in a cell with a pathlength L is given by the Beer‑Lambert law as:
(5)
where Bλ corresponds to the background absorbance of the sample and e is an error term due to instrumental noise. Provided that the values of the extinction coefficients: ελ (HI-) and ελ (I2-) have been measured as a function of wavelength, absorbance measurements made at two or more wavelengths can be used to estimate the ratio [I2-] / [HI-].
In the case that only two wavelengths are used, and provided that the background can be eliminated effectively by a subtractive procedure, Equation (5) can be rearranged to give (assuming no instrumental error)
(6)
where the numbers 1 and 2 refer to the wavelengths chosen. For the best sensitivity, the wavelengths corresponding to the absorbance maxima of the base (I2-) and acid (HI-) forms, respectively, are used. The various terms ε are the extinction coefficients of the specified species at wavelengths 1 and 2, respectively.
5.0 Reagents
Use only reagents of recognized analytical grade.
5.1 m‑cresol purple, containing no spectrophotometric impurities.
NOTE 1 Liu, X., Patsavas, M.C. and Byrne, R.H. (2011), Reference [13], showed the indicator can be characterized and purified using the HPLC system. The wavelength of isosbestic point for HI- / I2- of the pure m‑cresol purple,
λisos (HI- / I2-) depends on the following equation: λisos( HI- / I2- ) = 490,6 – 0,10 t, where t is the temperature in degrees Celsius. That for H2I / HI- is also λisos( H2I / HI- ) = 482,6 – 0,10 t.
NOTE 2 Liu, X., Patsavas, M.C. and Byrne, R.H. (2011), Reference [13], and Patsavas, M.C., Byrne, R.H. and Liu, X. (2013), Reference [16], describe the purification method of m‑cresol purple.
NOTE 3 It is highly recommended to use purified m‑cresol purple, however in the case it isn’t available, it is required to use the correction procedure described in Douglas, N.K., and Byrne, R.H. (2017), Reference [11].
5.2 Solution of pure m‑cresol purple
A concentrated (at least 2 mmol/l) pure indicator solution of known is prepared either in ultrapure water or in a 0.7M NaCl ionic background. The pH value of the dye solution is adjusted to be in the range 7,9 ± 0,1 pH units - chosen to match pHT measurements from an oceanic profile by adding HCl or NaOH as needed; this implies that for m‑cresol purple A1/A2 ≈ 1,6.
NOTE The absorbance ratio of a concentrated indicator solution can be measured using a cell with a short pathlength (0,5 mm).
5.3 Ultrapure water, of resistivity about 18 MΩ cm.
To be used in the preparation of the pure m‑cresol purple solution and in rinsing of equipment (e.g. cells).
6.0 Apparatus
Usual laboratory equipment and, in particular, the following:
6.1 Flexible drawing tube
Approximately, 40 cm long, sized to fit snugly over cell port. Silicone rubber is suitable. The drawing tube can be pre‑treated by soaking in clean sea water for at least one day. This minimizes the amount of bubble formation in the tube when drawing a sample.
6.1.1 Spectrophotometric cell
These should be made of optical glass with a 10 cm pathlength, two ports and stoppers. A sufficient number of cells are needed to collect all the samples that will be analysed from a particular cast.
NOTE A flow through cuvette with a 10 cm pathlength is also available. Sample bottles of at least 200 ml with air tight caps are needed to use the cuvette.
6.1.2 Micropipette
A micropipette is used to add the indicator to the cell. It should be of ~0,1 cm3 capacity with a narrow tube attached to act as a nozzle.
6.1.3 High‑quality spectrophotometer
For work of the highest sensitivity and precision, a double‑beam spectrophotometer is desirable. However, good results can be obtained with a high‑quality single‑beam instrument.
6.1.4 Temperature-control system for spectrophotometer cell
Commercially manufactured, thermostated spectrophotometer compartments that can accommodate 10 cm cells are rarely available and one will probably have to be custom‑made. Equipment need to be maintained in a close room with controlled temperature, and all the materials, solutions and samples need to be maintained in the same room for at least 2-h prior to start the test. The temperature should be regulated to within 0,1 °C.
6.1.5 System to warm samples to measurement temperature
Although it is possible to warm up the cells containing samples in the sealed bags in a thermostat bath, this is inconvenient. It is much better to build a custom‑made thermostated compartment that can hold approximately 12 cells at once without getting them wet.
6.1.6 Thermostat bath
A thermostat bath is used to control the temperature of both the cell compartment and the system to ±0,05 °C
7.0 Sampling
Sea water should be collected from the sampler bottle immediately after opening, and before a significant amount of water is removed, to ensure representative sampling. This is necessary to minimize exchange of CO2 with the air space in the sampler which affects all carbon parameters except total alkalinity. It is desirable that the carbon samples be collected before the sampler bottle is half empty and within 10 min of it being first opened.
Rinse the sample bottle — If the bottle is not already clean, rinse it twice with 30 cm3 to 50 cm3 of sample to remove any traces of a previous sample.
Fill the sample bottle — Fill the bottle smoothly from the bottom using a drawing tube which extends from the sampler bottle drain to the bottom of the sample bottle. It is critical to remove any bubbles from the draw tube before filling. Overflow the water by at least a full bottle volume. The air space within the sample bottle is kept to a minimum, References [9] and [10]. It is allowed to draw the sample directly from sampler bottle into the optical cell with two ports.
It is recommended that the pH analysis must be performed immediately after sampling, although storage experiments showed that the pH of sea water is stable up to 24 h even in the case of coastal water (see Annex B). However, while awaiting analysis, store the samples in the refrigerator or icebox (not frozen). The stability has not been assessed in case of important primary productivity (e.g. during a plankton bloom): in this case, it could be recommended to poison the seawater.
8.0 Procedure
In the case of optical cell with two ports, warm sample cell to 25,0 ± 0,1 °C by placing a number of cells in a thermostated compartment (see 6.6) for at least 1h. Clean and dry the exterior of the cell before placing it in the thermostated sample compartment of the spectrophotometer.
Measure and record the absorbances at three wavelengths: at the wavelengths corresponding to the absorption maxima of the base (I2-) and acid (HI-) forms of the m‑cresol purple, 578 nm and 434 nm, respectively, and a non‑absorbing wavelength (750 nm).
NOTE 1 750 nm is chosen as a non-absorbing wavelength because it lies in a flat absorbance region, making it relatively insensitive to wavelength accuracy. This allows for the correction of any baseline shifts that may occur during the measurement process.
Remove one of the cell caps to inject m‑cresol purple: add approximately 0,05 cm3 to 0,1 cm3 of m‑cresol purple to the sample. Replace the cap and shake the cell to mix the sea water and m‑cresol purple. Absorbance values between 0,4 and 1,0 at each of the two absorbance peaks should be obtained by adding the appropriate amount of m-cresol purple.
Return the cell to the spectrophotometer, again measure and record the absorbances at three wavelengths of sea water after adding m‑cresol purple.
Cells should be positioned to maintain consistent alignment(s) between two absorbance measurements.
NOTE 2 In the case of a flow-through cuvette, warm the pH sample bottles to 25,0 ± 0,1 °C. Measure and record the reference absorbances at three wavelengths as described above. Then, add m-cresol purple to the sea water, gently mix the sample, and immediately measure and record the absorbances at three wavelengths, ensuring that the absorbance values fall between 0,4 and 1,0 at both absorbance peaks.
For routine samples, a single measurement is sufficient. However, multiple measurements of a known reference material are recommended to verify both the precision and accuracy of the method.
9.0 Calculation and expression of results
9.1 Correction of measured absorbances
At each of the three wavelengths, subtract the absorbances measured for the background measurement (without indicator) from the corresponding absorbances measured for the system containing indicator.
In addition, the absorbance measured at a non‑absorbing wavelength is used to monitor and correct for any baseline shift due to error in repositioning the cell, instrumental shifts, etc. This assumes that the magnitude of any observed baseline shift is identical across the visible spectrum. Subtract the measured shift from the background-corrected absorbances at wavelengths 1 and 2 to obtain the final corrected absorbance value at each wavelength. These final absorbance values, corrected for background absorbances and any observed baseline shifts, are used to calculate A1/A2, the absorbance ratio which describes the extent of protonation of the indicator.
The difference between the baseline absorbance (sea water only) and the absorbance of the sample and indicator at 750 nm should be no greater than ±0,001. If this value is exceeded, the cell should be removed and the optical windows cleaned before the absorbances are measured again.
9.1.1 Calculation of the pHT of the sea water and indicator
The pHT of the sea water and indicator in the cell is computed from
(7)
Equation (7) can be rewritten as:
(8)
where K2 is the equilibrium constant between the species HI- and I2- (expressed on the total hydrogen ion concentration scale in mol/kg), and A1 and A2 are the corrected absorbances measured at the wavelengths corresponding to the absorbance of 578 nm and 434 nm, respectively.
The first term of Equation (8), -log10 { K2ε1 (I2-) /ε2(HI-) }, is a function of salinity and temperature and has been determined by careful laboratory measurements.
-log10 { K2ε1 (I2-) / ε2(HI-) } = a + b / T + c lnT - dT
where parameters of a, b, c and d are in the range of 278,15 ≤ T ≤ 308,15 K and 20 ≤ S ≤ 40.
a = -246,642 09 + 0,315 971 S + 2,885 5 × 10-4 S2
b = 722 9,238 64 – 7,098 137 S – 0,057 034 S2
c = 44,493 382 – 0,052 711 S
d = 0,0781 344
The various extinction coefficient terms ε correspond to values measured for the specified species at wavelengths 578 nm and 434 nm and are defined as follows:
ε1(HI-) / ε2(HI-) = -0,007 762 + 4,517 4 × 10-5T
ε2(I2-) / ε1(I2-) = -0,020 813 + 2,602 62 × 10-4T + 1,043 6 × 10-4 ( S -35 )
NOTE The -log10 { K2ε1(I2-) / ε2(HI-) } given here and the various extinction coefficient terms ε are those from Liu, X., Patsavas, M.C. and Byrne, R.H. (2011), Reference [13].
9.1.2 Correction for pHT change resulting from addition of the indicator
The addition of indicator to the sea water sample will perturb the pHT (another acid–base system has been added). Although care is taken to minimize this (by adjusting the indicator solution pH), it is desirable to correct for the addition of indicator to obtain the best pHT measurements.
Although, in principle, the pHT perturbation could be calculated from knowledge of the equilibrium chemistry of the sample and the indicator, it is simpler to evaluate the magnitude of the correction empirically. A pair of additions of indicator is made to each of a series of sea water samples with different pHT values, and the change in the measured ratio (A1/A2) with the second addition of indicator solution is determined as a function of the measured value (A1/A2) determined after the first addition of indicator using a least squares procedure.
(9)
where V is the volume of indicator added at each addition. The final, corrected, absorbance ratio is given in Equation (10):
(A1/A2)corr = (A1/A2) – V [a + b (A1/A2)] (10)
NOTE Absorbance at the isosbestic point wavelength for HI-/I2- depends on the concentration of indicator. Then, (A1/A2)corr can be determined by the linear regression of (A1/A2) and the absorbance at the isosbestic point wavelength.
Another method is to plot the ratio (A1/A2) as a function of the volume of dye added. This allows, by a linear extrapolation to a volume equal to zero, to compute the corrected (A1/A2) ratio to be used for the calculation of the pHT.
If exactly the same volume of indicator was added in the first and second addition, the corrected (A1/A2) ratio can be calculated with the following equation:
(A1/A2) corrected = 2 (A1/A2) first_addition - (A1/A2) second_addition (11)
For sea water samples with similar salinity and pH, the correction can be calculated for one sample using the double or triple dye addition method described above. This same correction can then be applied to the other measurements
(informative)
Performance data- Interlaboratory comparison on TRIS buffer and natural sea water
This interlaboratory comparison (ILC) exercise has been proposed to partners and stakeholders of the marine community, in the framework of the EURAMET SapHTies project, in autumn 2023.
The objectives pursued by the ILC were the following:
— validating the capacity of the improved spectrophotometric pHT method to meet the needs necessary for the quantification of ocean acidification and the capacity of laboratories to apply these new recommendations. The targeted pHT expanded uncertainty is 0,006.
— proposing improvements of the current QA/QC practices implemented in the different oceanographic observation networks and target a possible completion of the ISO18191:2015 “Water quality — Determination of pHT in sea water — Method using the indicator dye m‑cresol purple” standard.
- ILC samples
Two samples were selected for the realisation of this ILC:
1. A TRIS buffer made of a salted matrix was proposed as the best standardised current reference material to represent a sea water matrix. This commonly accepted reference material for sea water pHT is an equimolal buffer of 2-Amino-2-(hydroxymethyl)-1,3-propanediol (TRIS) and TRIS hydrochloride (TRIS.HCl) in artificial sea water of salinity 35 (Wolfe et al., 2021, reference [21]). This buffer can be certified in terms of pHT by potentiometric measurement made in an electrochemical cell without junction called Harned cell. It is provided in 160 ml bottles.
2. A natural sea water was also included in the ILC to collect measurement similar to the one produced in routine applications. This sample from the OSIL Company has been selected for its specific production features. The sample is free from any biological content that can disturb its stability in pHT. It is not added with any chemical materials that might interact with the chemicals used for spectrophotometric pHT measurements. The sample is supplied in 250 ml borosilicate bottles with limited air headspace, reducing possible pHT evolution coming from equilibrium with air CO2 content. Moreover, this sample has been surveyed in several carbonates studies and a preliminary test performed for the objectives of this ILC indicated promising batch homogeneity.
- Dye
A purified meta-cresol purple (m-CP) solution of 2 mM, dissolved in a basic solution (NaOH in 0,7 mol/kg NaCl) and adjusted to pH~8 was provided to the participants in the framework of the ILC. Indeed, the dye used when performing pHT measurement is entirely correlated to the measurement obtained. It was thus crucial to provide to all, exactly the same product. This dye was specifically made by GEOMAR for the purposes of the SapHTies project and fully characterized during the lifetime of the project; its extinction coefficients were determined as a function of temperature and salinity. This dye has been produced in one batch, then aliquoted in 12 ml amber vials.
- Measurement protocol
The measurement protocol has been thoroughly described in a document distributed to the participants. Special attention was drawn on instruments and sample conditioning, sample sampling, temperature conditioning and avoiding sample pollution. The measurement itself was detailed in terms of measurement wavelengths, blanks and corrections to take care of, and a specific protocol for correcting for the dye addition in the sample has been imposed. This document was accompanied by a video to ease the realisation of the measurements: https://embed.ifremer.fr/videos/b3596a386e6a49429a4e5dd7fa33bd1b
A reporting table has been sent to each participant, requesting two types of measurements:
— Raw absorbance and temperature measurements, to assess the performances of the spectrophotometric method
— pHT measurements at the measurement temperature and pHT measurements reported to 25 °C, to evaluate the ability of each laboratory to estimate the pHT.
To ensure reliable statistical analysis, participants were asked to return at least 5 individual measurements each.
- Assessment of the reference value of TRIS buffer
The material has been characterised with Harned cells with a reference value of 8,096 0 ± 0,005 4 (k = 2; 25 °C) for a shell life of 2 months. Homogeneity tests, stability tests as well as the evaluation of the total uncertainty budget were performed according to (Capitaine et al. 2023, reference [6]).
- Homogeneity, stability and metrological traceability of natural sea water
Several bottles from the production batch of natural sea water included in the ILC, have been tested for homogeneity and stability. Concerning homogeneity, it should be noted that spectrophotometric measurements, due to their own repeatability, are limited to demonstrate the homogeneity of the batch. Taking in account the measurement variations that can be observed within one bottle, the material was considered homogeneous for the purpose of the ILC.
The stability of the batch has been surveyed for several months, before, over and after the course of the ILC. Despite the variability of measurements, a global trend of increasing pHT was observed. This increase was less than 0,001 pHT unit over the 2 weeks of the ILC.
Natural sea water pHT measurement cannot currently qualify for any recognised way of traceability to the International Standard of Unit (SI). Several limitations are part of this lack of traceability, one of them being the technological inapplicability of the recognized Harned cell primary method on sea water matrix.
- Assessment of the pHT spectrophotometric method performance according to NF ISO 5725-2
In order to assess the performances of the practical realisation of the pHT spectrophotometric method itself, a unique post-processing was applied on raw data reported by laboratories (raw absorbances together with temperature measurements). In this way, variations due to modelling were reduced to a minimum.
- TRIS buffer
The Cochran test indicated no isolated laboratory. The Grubbs test performed on laboratories averages indicated no isolated laboratory. The statistical analysis of all results collected for the TRIS buffer is presented in Table A.1.
- Natural sea water
The Cochran test indicated no isolated laboratory. The Grubbs test performed on laboratories averages indicated no isolated laboratory. The statistical analysis of all results collected for the natural sea water sample is presented in Table A.1.
Table A.1 — Performance data for TRIS buffer and natural sea water
Sample | Matrix | l | n | o % | sR | CV,R % | sr | CV,r % | |
1 | TRIS | 12 | 80 | 0 | 8,092 3 | 0,005 3 | 0,066 | 0,004 1 | 0,051 |
2 | sea water | 12 | 97 | 0 | 7,899 6 | 0,003 9 | 0,049 | 0,002 9 | 0,036 |
l number of laboratories after outlier rejection n number of analytical results after outlier rejection o percentage of outliers
sR reproducibility standard deviation CV,R coefficient of variation of reproducibility sr repeatability standard deviation CV,r coefficient of variation of repeatability |
- Interlaboratory comparison on North Pacific sea water
An interlaboratory trial was organised by Dr. Shuichi Watanabe and supervised by Dr. Koh Harada, Convenor of ISO/TC 147/SC 2/WG 67 with the assistance of Dr. Yoshiyuki Nakano and Mr. Koichi Goto and was performed in spring 2014. A total of 11 laboratories from 8 countries participated (Japan: 3, France: 2, Canada: 1, Korea: 1, Norway: 1, Portugal: 1, Sweden: 1, UK: 1). Two participants did not adhere to the method description. Their results were excluded from the statistical evaluation of the interlaboratory trial.
Five bottles of 500 ml sea water sample which were subsampled from one large container were sent to each laboratory. The sea water was collected in the western North Pacific and its salinity was 34,399.
The results are given in Table A.2.
Table A.2 — Performance data for sea water pH
Sample | Matrix | l | n | o % | sR | CV,R % | sr | CV,r % | |
1 | Sea water | 8 | 107 | 11,1 | 7,992 | 0,003 | 0,040 | 0,002 | 0,030 |
l number of laboratories after outlier rejection n number of analytical results after outlier rejection o percentage of outliers
sR reproducibility standard deviation CV,R coefficient of variation of reproducibility sr repeatability standard deviation CV,r coefficient of variation of repeatability |
(informative)
Storage stability
The storage stability of sea water sample for pHT was confirmed by sequential experimentation over 24 h. Sea water samples were collected at the coastal sea of Aomori Prefecture, Japan by a plastic bucket. Salinity of the sea water was 33,532. Sea water samples 1 were sampled in 250 ml glass bottles from a bucket. Samples 2 were filtered through 25 mm-diameter Whatman GF/F filter in laboratory and sampled in 250 ml glass bottles. Samples 3 were sampled in 250 ml glass bottles from a bucket and poisoned with 100 μl of over saturated solution of mercury chloride. These samples were stored in a refrigerator and got out an hour and a half before analysis.
The results are given in Table B.1.
Table B.1 — Storage stability of sea water pHT
Sample | Matrix | Date Time dd hh:mm | pHT at 25 °C |
1 | Sea water | 16 13:29 | 7,953 |
1 | Sea water | 16 13:48 | 7,955 |
1 | Sea water | 16 16:07 | 7,953 |
1 | Sea water | 16 18:14 | 7,952 |
1 | Sea water | 17 09:55 | 7,950 |
1 | Sea water | 17 12:40 | 7,952 |
1 | Sea water | 17 15:03 | 7,953 |
Average | 7,952 | ||
Standard deviation | 0,001 | ||
2 | Filtrated sea water | 16 16:24 | 7,941 |
2 | Filtrated sea water | 16 16:32 | 7,942 |
2 | Filtrated sea water | 16 18:32 | 7,941 |
2 | Filtrated sea water | 17 10:12 | 7,939 |
2 | Filtrated sea water | 17 12:58 | 7,932 |
2 | Filtrated sea water | 17 15:22 | 7,938 |
2 | Filtrated sea water | 17 17:34 | 7,940 |
Average | 7,939 | ||
Standard deviation | 0,003 | ||
3 | Sea water + HgCl2 | 16 13:40 | 7,953 |
3 | Sea water + HgCl2 | 16 13:56 | 7,952 |
3 | Sea water + HgCl2 | 16 16:15 | 7,956 |
3 | Sea water + HgCl2 | 16 18:24 | 7,956 |
3 | Sea water + HgCl2 | 17 10:04 | 7,955 |
3 | Sea water + HgCl2 | 17 12:48 | 7,956 |
3 | Sea water + HgCl2 | 17 15:14 | 7,953 |
Average | 7,955 | ||
Standard deviation | 0,002 |
Bibliography
[1] AßMANN, S., FRANK, C. and KÖRTZINGER, A. (2011) Spectrophotometric high-precision seawater pH determination for use in underway measuring systems. Ocean Sci., 7, pp. 597-607.
[2] Bellerby, R.G.J., Olsen, A., Johannessen, T. and Croot, P. (2002) A high precision spectrophotometric method for on-line shipboard seawater pH measurements: the automated marine pH sensor (AMpS). Talanta 56 (1), pp. 61-69.
[3] Byrne, R.H., Robert-Baldo, G., Thompson, S.W. and Chen, C.T.A. (1988) Seawater pH measurements: an at-sea comparison of spectrophotometric and potentiometric methods. Deep-Sea Res., 35, pp. 1405–1410.
[4] Byrne, R.H. and Breland, J.A. (1989) High precision multiwavelength pH determinations in seawater using cresol red. Deep-Sea Res., 36, pp. 803–810.
[5] Byrne, R.H., Liu, X., Kaltenbacher, E. and Sell, K. (2002) Spectrophotometric measurement of total inorganic carbon in aqueous solutions using a liquid core waveguide. Analytica Chimica Acta, 451(2): pp. 221-229.
[6] Capitaine, G., Stoica, D., Wagener, T., Fisicaro, P. (2023) Production of a reference material for seawater pHT measurements by a National Metrology Institute. Marine Chemistry 252, 104244.
[7] Clayton, T.D. and Byrne, R.H. (1993) Spectrophotometric seawater pH measurements: total hydrogen ion concentration scale calibration of m-cresol purple and at-sea results. Deep-Sea Res., 40, pp. 2115‑2129.
[8] Dickson, A.G. (1993) The measurement of sea water pH. Marine Chem., 44, pp. 131–142.
[9] Dickson, A.G. (1993) pH buffers for sea water media based on the total hydrogen ion concentration scale. Deep-Sea Res., 40, pp. 107–118.
[10] Dickson, A.G., Sabine, C.L. and Christian, J.R. (Eds.) (2007) Guide to best practices for ocean CO2 measurements. PICES Special Publication 3, 191 pp.
[11] Douglas, N.K., and Byrne, R.H. (2017) Achieving accurate spectrophotometric pH measurements using unpurified meta-cresol purple. Marine Chemistry, 190, 66–72
[12] Karl M.D., Winn D.C., Hebel V.W. D and LETELIER, R. (1990) Hawaii ocean time-series program field and laboratory protohttp://hahana.soestcols. .hawaii.edu/hot/protocols/chapi.html.
[13] Liu, X., Patsavas, M.C. and Byrne, R.H. (2011) Purification and characterization of meta-cresol purple for spectrophoto-metric seawater pH measurements. Environ. Sci. Technol., 45, pp. 4862-4868.
[14] Liu, X., Wang, Z.A., Byrne, R.H., Kaltenbacher, E.A., and Bernstein, R.E. (2006) Spectrophotometric measurements of pH in-situ laboratory and field. Evaluations of Instrumental Performance. Environ. Sci. Technol., 40, pp. 5036-5044.
[15] Nakano, Y., Kimoto, H., Watanabe, S., Harada, K. and Watanabe, Y.W. (2006) Simultaneous measurements of in situ pH and CO2 in the sea using spectrophotometric profilers. Journal of Oceanography, 62, pp. 71‑81.
[16] Patsavas, M.C., Byrne, R.H. and Liu, X. (2013) Purification of meta-cresol purple and cresol red by flash chromatography: Procedures for ensuring accurate spectrophotometric seawater pH measurements. Mer. Chem., 150, pp. 19-24.
[17] Easley, R.A. and Byrne, R.H. (2012) Spectrophotometric calibration of pH electrodes in seawater using purified m-cresol purple. Environ. Sci. Technol., 46, pp. 5018-5024.
[18] Tapp, M., Hunter, K., Currie, K. and Mackaskill, B. (2000) Apparatus for continuous-flow underway spectrophotometric measurement of surface water pH. Mar. Chem., 72(2-4), pp. 193-202.
[19] Seidel, M.P., Degrandpre, M.D., and Dickson, A.G. (2008) A sensor for in situ indicator-based measurements of seawater pH. Mar. Chem., 109, pp. 18-28.
[20] Wang, Z.A., et al. (2007) Simultaneous spectrophotometric flow-through measurements of pH, carbon dioxide fugacity, and total inorganic carbon in seawater. Anal. Chim. Acta 596 (1), pp. 23-36.
[21] Wolfe, W.H., Shipley, K.M., Bresnahan, P.J., Takeshita, Y., Wirth, T., Martz, T.R. (2021) Technical note: Stability of tris pH buffer in artificial seawater stored in bags. Ocean Sci. 17, 819–831.
[22] Yao, W., Lui, X. and Byrne, R.H. (2007) Impurities in indicators used for spectrophotometric seawater pH measurements: Assessment and remedies. Mer. Chem., 107, pp. 167-172.