ISO/DIS 4259-2
ISO/TC 28
Secretariat: NEN
Date: 2025-02-18
Petroleum and related products — Precision of measurement methods and results —
Part 2:
Interpretation and application of precision data in relation to methods of test
Produits pétroliers et connexes — Fidélité des méthodes de mesure et de leurs résultats —
Partie 2: Application des valeurs de fidélité relatives aux méthodes d'essai
DIS stage
Warning for WD’s and CD’s
This document is not an ISO International Standard. It is distributed for review and comment. It is subject to change without notice and may not be referred to as an International Standard.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to provide supporting documentation.
© ISO 2025
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: + 41 22 749 01 11
E-mail: copyright@iso.org
Website: www.iso.org
Published in Switzerland
Contents
4 Application and significance of repeatability, r, and reproducibility, R 2
6 Assessment of quality conformance to specification 7
6.2 Assessment of quality conformance by the supplier 8
6.3 Assessment of quality conformance by the recipient 9
7.1 Resolve dispute by negotiation 11
7.2 Use of the test method or procedure in case of dispute 11
7.3 Dispute resolution procedure 12
7.5 Example of a dispute resolution 13
Annex A (informative) Explanation of formulae given in Clause 4 15
Annex C (informative) General approach to bias assessment using multiple materials 20
Annex D (informative) Glossary 21
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO document should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 28, Petroleum and related products, fuels and lubricants from natural or synthetic sources.
This second edition cancels and replaces the first edition (ISO 4259-2:2017), which has been technically revised.
The main changes are as follows:
- included normative references to ISO 4259-3 and ISO 4259-4;
- deleted annex C because its content is covered by ISO 4259-4;
- improved Figures 2 and 3;
A list of all parts in the ISO 4259 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.
For purposes of setting product specifications, and to check product compliance against these specifications, standard test methods are usually referenced for specific properties of commercial petroleum and related products. Two or more measurements of the same property of a specific sample by a specific test method, or by different test methods that purport to measure the same property, will not usually give exactly the same result. It is, therefore, necessary to take proper account of this fact when setting product specifications, assessing if the differences between test results are within statistical expectation, and making specification compliance decisions based on limited test results. By using statistically-based estimates of the precision for a test method, the following can be achieved:
- an objective measure of the reliability of specification limits,
- a specification compliance decision, and
- the degree of agreement expected between two or more results obtained in specified circumstances.
This document describes the applications of the precision of test method as derived from ISO 4259-1. It is intended to be a companion document to ISO 4259-1. Additional normative and informative discussions on how to use this precision to assess the “in statistical control” status and precision capability of a specific laboratory in the execution of a test method are provided. Also, the general approach to the agreement between two different test methods that purport to measure the same property is given.
ISO 4259-1 and ISO 4259-2 encompass both the determination of precision estimates and the application of precision data. It attempts to be aligned with ASTM D6300[1] regarding the determination of the precision estimates and with ASTM D3244[2] for the utilization of test data.
A glossary of the variables used in this document and ISO 4259-1 is included in ISO 4259-1:2024, Annex I.
Petroleum and related products — Precision of measurement methods and results —
Part 2:
Interpretation and application of precision data in relation to methods of test
1.0 Scope
This document specifies the methodology for the application of precision estimates of a test method derived from ISO 4259-1. In particular, it defines the procedures for setting the property specification limits based upon test method precision where the property is determined using a specific test method, and in determining the specification conformance status when there are conflicting results between supplier and receiver. Other applications of this test method precision are briefly described in principle without the associated procedures.
The procedures in this document have been designed specifically for petroleum and petroleum-related products, which are normally homogeneous. However, the procedures described in this document can also be applied to other types of homogeneous products. Careful investigations are necessary before applying this document to products for which the assumption of homogeneity can be questioned.
2.0 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 4259-1, Petroleum and related products — Precision of measurement methods and results — Part 1: Determination of precision data in relation to methods of test
ISO 4259-3, Petroleum and related products — Precision of measurement methods and results — Part 3: Monitoring and verification of published precision data in relation to methods of test
ISO 4259-4, Petroleum and related products — Precision of measurement methods and results — Part 4: Use of statistical control charts to validate 'in-statistical-control' status for the execution of a standard test method in a single laboratory
3.0 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 4259-1 and the following apply.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https://www.iso.org/obp
- IEC Electropedia: available at https://www.electropedia.org/
proficiency testing scheme
PTS
program designed for the periodic evaluation of participating laboratories’ testing capability of a standard test method through the statistical analysis of their test results obtained on aliquots prepared from a single batch of homogeneous material
Note 1 to entry: The frequency of such testing varies in accordance with the program objective. Each execution of testing involves testing of a single batch of material. Materials typically vary from test to test.
Note 2 to entry: This is also commonly referred to as Inter Laboratory Cross Check Program (ILCP).
recipient
individual or organization who receives or accepts a product delivered by the supplier (3.3)
supplier
individual or organization responsible for the quality of a product just before it is taken over by the recipient (3.2)
4.0 Application and significance of repeatability, r, and reproducibility, R
4.1 General
The value of these quantities is estimated from analysis of variance (two-factor with replication) performed on the results obtained in a statistically designed inter-laboratory programme in which different laboratories each test a range of samples. Repeatability and reproducibility values estimated in accordance with ISO 4259-1 or other statistical techniques shall be included in each published test method.
NOTE See Annex A for an account of the statistical reasoning underlying the formulae Clause 4.
In the following clauses, it is assumed that the result(s) are obtained from a test method that is in statistical control. For determination of “in statistical control”, see ISO 4259-4.
4.1.1 Repeatability, r
4.1.2 General
Most laboratories do not carry out more than one test on each sample for routine quality control purposes except in some circumstances, such as in cases of dispute or if the test operator wishes to confirm that his technique is satisfactory. In such circumstances, when multiple results are obtained, it is useful to check the consistency of repeated results against the repeatability of the method. The appropriate procedure is outlined in 4.2.2. It is also useful to know what degree of confidence can be placed on the average results, and the method of determining this is given in 4.2.3.
4.1.3 Acceptability of results
When only two results are obtained under repeatability conditions and their difference is less than or equal to r, the test operator may consider his work as being under control and may take the average of the two results as the estimated value of the property being tested.
If the two results differ by more than r, both shall be considered as suspect and at least three more results obtained. Including the first two, the difference between the most divergent result and the average of the remainder shall then be calculated and this difference compared with a new value, r1, instead of r, given in Formula (1):
(1)
where k is the total number of results obtained.
If the difference is less than or equal to r1, all the results shall be accepted. If the difference exceeds r1, the most divergent result shall be rejected and the procedure specified in this section repeated until an acceptable set of results is obtained.
The average of the acceptable results shall be taken as the estimated value of the property. However, if two or more results from a total of not more than 20 have been rejected, the operating procedure and the apparatus shall be checked and a new series of tests made, if possible.
4.1.4 Confidence limits calculations using results collected under repeatability conditions
When a single test operator, who is working within the precision limits of the method, obtains a series of k results under repeatability conditions, giving an average, , and the results meet the repeatability requirement in 4.2.2, it can be assumed with 95 % confidence that the true value, μ, of the characteristic lies within the following limits:
where
(3)
When k = 1, use the single test result as the value for the term as follows:
where R is the published test method reproducibility as discussed in 4.3.
Similarly, for the single limit situation, when only one limit is fixed (upper or lower), it can be assumed with 95 % confidence that the true value, μ, of the characteristic is limited as follows:
or
The factor 0,59 is the ratio , where 0,84 is derived in Annex A.
When r is much smaller than R, little improvement in the precision of the average is obtained by carrying out multiple testing under repeatability conditions.
4.2 Reproducibility, R
4.2.1 Acceptability of results
The procedure specified in 4.3 is intended for judging the acceptability, with respect to the reproducibility of the test method, of results obtained by different laboratories in normal, day-to-day operations and transactions. In cases of dispute between a supplier and a recipient, the procedure specified in Clauses 5 to 7 shall be adopted.
When single results are obtained in two laboratories and their difference is less than or equal to R, the two results shall be considered as acceptable, and used to calculate the average . The average , rather than either single result separately, shall be used as the estimated value of the tested property.
The true value μ of the characteristic is contained within the following limits with a 95 % confidence:
Similarly for the single limit situation, when only one limit is fixed (upper or lower), the true value μ of the characteristic is contained with the following limits with a 95 % confidence:
or
The factor 0,42 is the ratio of as it is an average of two results.
If the two results differ by more than R, both shall be considered as suspect. Each laboratory shall then obtain at least three other acceptable results (see 4.2.2).
In this case, the difference between the averages of all acceptable results of each laboratory shall be judged for conformity using a new value, R2, instead of R, as given by Formula (10):
(10)
where
| R | is the reproducibility of the method; |
| r | is the repeatability of the method; |
| k1 | is the number of results of the first laboratory; |
| k2 | is the number of results of the second laboratory. |
If the difference between the averages is less than or equal to R2, then these averages are acceptable and their overall average shall be considered as the estimated value of the tested property. If the difference between the averages is greater than R2, and there is a dispute on the specification conformance of the tested property, then the procedure specified in Clause 7 shall be adopted.
If circumstances arise in which there are more than two laboratories, each supplying one or more acceptable results, the difference between the most divergent laboratory average and the average of the remaining N laboratory averages shall be compared to R3:
where
(11)
(12)
R1 is given in Formula (3), and corresponds to the most divergent laboratory average.
If this difference is equal to or less than R3 in absolute value, all results shall be regarded as acceptable and their average taken as the estimated value of the property.
If the difference is greater than R3, the most divergent laboratory average shall be rejected and the comparison using Formulae (11) and (12) repeated until an acceptable set of laboratory averages is obtained. The average of these laboratory averages shall be taken as the estimated value of the property. However, if two or more laboratory averages from a total of not more than 20 have been rejected, the operating procedure and the apparatus shall be checked and a new series of tests made, if possible.
4.2.2 Confidence limits calculations using results collected under reproducibility conditions
When N laboratories obtain one or more results under conditions of repeatability and reproducibility, giving an average of laboratory averages , the true value μ of the characteristic is contained within the following limits with 95 % confidence:
(13)
Similarly for the single limit situation, when only one limit is fixed (upper or lower), the true value μ of the characteristic is contained with the following limits with 95 % confidence:
or
4.3 Use of reproducibility to determine bias between two different test methods that purport to measure the same property
4.3.1 General
For the situation where two different test methods purport to measure the same property, the reproducibility estimates (R) from the respective test methods shall be used in conjunction with the averages obtained from multiple laboratories for the same material to determine if a bias correction can be applied to improve statistically the agreement between the two methods for that material. For example, results collected through proficiency testing schemes (PTS) for different test methods using the same sample can be analysed in this fashion.
NOTE Discussion on methodology for this type of assessment for the simultaneous analysis of multiple materials / property levels that span the intersecting scope of two different test methods is beyond the scope of this document. Interested readers are encouraged to consult ISO 4259-5[3] for a detailed presentation on the subject. Annex C provides a brief overview on the general statistical approach for the aforementioned situation.
4.3.2 Process
Assume that Test Method A and Test Method B are test methods that purport to measure property C.
Calculate the following statistic:
where
| is the average from LA results for property C for a material using Test Method A, where each result is a single result obtained under reproducibility conditions; | |
| LA | is the total number of laboratories (results) for Test Method A and should be >20; |
| RA | is the reproducibility of Test Method A; |
| is the average from LB results for property C for a material using Test Method B on the same material tested by Test Method A, where each result is a single result obtained under reproducibility conditions; | |
| LB | is the total number of laboratories (results) for Test Method B, and should be >20; |
| RB | is the reproducibility of Test Method B. |
If Z > 2, it shall be concluded, with 95 % confidence, that a constant bias correction statistically improves the degree of agreement between Test Method A and Test Method B for property C for this material.
5.0 Specifications
5.1 Aim of specifications
The purpose of a specification is to specify an acceptable limit or limits to the true value, μ, of the property as determined by a specified test method. In practice, however, this true value can never be established exactly since the results obtained by applying the specified test method in a single or multiple laboratories can show acceptable scattering as defined by the repeatability and reproducibility. There is, therefore, always some uncertainty as to the true value of the tested property determined from a finite number of test results.
Petroleum product compliance with specifications is assessed in accordance with Clauses 6 and 7. By prior agreement a supplier and recipient may use the alternative procedures described in Annex B.
It is important that the test method specified for the determination of the property governed by the specification limit(s) is sufficiently precise to reliably determine whether or not the product meets the specifications.
5.1.1 Construction of specifications limits in relation to scope and precision of the specified test method
The specification limits shall not be outside the method scope limits as determined in ISO 4259-1.
The lower specification limit shall not be less than the lower scope limit of the test method, and the upper specification limit shall not be greater than the upper scope limit of the test method (see ISO 4259-1:2024; 6.5).
In addition, the distance between lower and upper specification limit shall also satisfy the following condition: upper specification limit minus lower specification limit shall not be less than the quantity 2R evaluated at lower method scope limit plus 2R evaluated at upper method scope limit. See Figure 1 for an illustration of this concept.
Usually, specifications deal with limits for the values of the properties. To avoid uncertainty, such limits are normally expressed as “not less than” or “not greater than”. Limits are of two types:
- a double limit, upper and lower, for example viscosity not less than 5 mm2/s and not greater than 16 mm2/s; boiling point 100 °C ± 0,5 °C;
- a single limit, upper or lower, for example water content not greater than 2 %; sulfur content not greater than 10 mg/kg ; solubility of bitumen not less than 99 %.
Figure 1 — Specification setting
In cases where, for practical reasons, the value of (A1 − A2) is less than the above minimum range requirement in Figure 1, the results obtained will be of doubtful significance in determining whether a sample does or does not satisfy the requirements of the specification. According to statistical reasoning, it is desirable for (A1 − A2) to be considerably greater than the above minimum range requirement. If not, one or both of the following courses shall be adopted:
- the specification limits shall be examined to see whether they can be widened to fit in with the precision of the test method;
- the test method shall be examined to see whether the precision can be improved, or an alternative test method adopted with an improved precision, to fit in with the desired specification limits.
Conformity to this document requires specifications to be drawn up in accordance with the above principles.
6.0 Assessment of quality conformance to specification
6.1 General
Clause 6 provides general information to allow the supplier or the recipient to judge the quality of a product with regard to the specification based on a single test result as obtained by the supplier or recipient. If both the supplier's and recipient's single test results are available, the estimate of the true value shall be obtained in accordance with 4.3. If the recipient decides to dispute the quality conformance to specification after examining his single result, or the estimate from 4.3, the procedure specified in Clause 7 shall be adopted.
As a prerequisite for acceptance for laboratory test results to be used in 6.2, 6.3, Clause 7 and Annex B, the conditions in 6.1.2 to 6.1.4 shall be satisfied.
Each laboratory's test result shall be obtained from a test method that is in statistical control in terms of precision and bias, as substantiated by in-house SQC charts following the methodologies described in ISO 4259-4 or other equivalent statistical techniques.
The standard deviation from the control charts (or equivalent statistical techniques) in 6.1.2, as calculated from at least 30 most recent results obtained over at least 15 days, with results that are separated by at least 8 h, shall not exceed the published test method standard deviation (R / 2,77).
If evidence exists from the published results of multiple PTS, that the R for a published test method is statistically inconsistent with the R actually achieved, the latter may be used in lieu of the published R to judge conformance to this clause, provided all of the following conditions are met:
- if legally permissible, and
- if the R calculated from multiple PTS have sufficient degrees of freedom (>30) using results that have been properly screened for outliers in accordance with GESD protocol in ISO 4259-1 or other equivalent statistical technique, and
- upon mutual agreement between disputing parties.
Each laboratory shall be able to demonstrate, by way of results from participation in PTSs, if available, a sustained testing proficiency and a lack of bias relative to PTS averages assigned in accordance with ISO 4259-3 Annex B or equivalent statistical techniques for the appropriate test method(s). In the event that a suitable PTS is not available, proficiency shall be demonstrated by way of testing certified reference materials (CRM) and in-house control charts on quality control (QC) samples, or by other method validation techniques acceptable to both parties.
6.1.1 Assessment of quality conformance by the supplier
A supplier who has no other source of information on the true value of a characteristic than a single result shall consider, with 95 % confidence, that the product meets the specification limit, only if the result, Xs, is such that
- in the case of a single upper limit, A1:
- in the case of a single lower limit, A2:
- in the case of a double limit (A1 and A2), both these conditions are satisfied (see 4.2.3).
The 95 % confidence decision limits as calculated using Formulae (17) and (18) are for the guidance of the supplier, and are not to be interpreted as an obligation. A reported value between the specification value and the limit from Formula (17) or (18) is not proof of non-compliance, but is an indication that the confidence for the product to meet the specification limit is less than 95 %. If the result is exactly at the specification limit, the probability of a re-test result meeting specification, by either the supplier or the recipient, is 50 %. A direct consequence of releasing product with a low confidence is that the probability of the receiver obtaining an off-specification result will be high.
The supplier shall only release the product if their test result meets specification or by mutual agreement with the customer.
If multiple results are obtained by the supplier under repeatability conditions, the average of the acceptable results and R1 as determined in 4.2.2 and 4.2.3 shall be used by the supplier as the basis for determination of specification conformance.
In the event of a dispute with the recipient, procedures in Clause 7 shall be followed.
6.1.2 Assessment of quality conformance by the recipient
6.1.3 General
Figure 2 is a procedure flowchart that describes steps outlined in this subclause, taking into account all available data and requirements of this document.
6.1.4 Single batch of product
A recipient who has no other source of information on the true value of a characteristic than a single result shall consider that the product fails the specification limit with 95 % confidence, only if the result, x, is such that:
- in the case of an upper limit of the specification A1,
- in the case of a lower limit of the specification A2,
- in the case of a double limit (A1 and A2), either of these conditions applies.
The 95 % confidence decision limits as calculated using Formulae (19) and (20) is for the guidance of the recipient, and is not to be interpreted as an obligation. A reported value between the specification value and the limit from Formula (19) or (20) is not proof of non-conformance, but is an indication that the confidence for the product failing specification limit is less than 95 %. If the result is exactly at the specification limit, the probability of a re-test meeting specification, by either the supplier or recipient, is 50 %.
If multiple results are obtained by the recipient under repeatability conditions, the average of the acceptable results and R1 as determined in 4.2.3 shall be used as the basis for determination of specification conformance.
If the recipient decides to dispute the specification conformance status for the batch in question, regardless of the recipient's result that was used as a basis for the decision to dispute, the procedures in Clause 7 shall be adhered to.
6.1.5 Multiple batches of product
Persistent results over multiple batches that fail to meet the specification limit, but by an amount not greater than 0,59R, is a strong indication that the product release confidence by the supplier is less than 95 %. If the latter is not acceptable by the recipient, it is recommended that the recipient contact the supplier and arrive at a mutually satisfactory resolution.
NOTE Five results in a row that fail to meet the specification limit constitute compelling evidence (greater than 95 % confidence) that at least one of the batches does not meet specification.
Figure 2 — Flowchart for assessment of specification conformance by recipient
6.1.6 Procedure for recipient to assess conformance for a single batch of product
In the case of assessing the conformance to specification of a single batch of product, the following example shows how to evaluate the process. The Research Octane Number (RON) specification within EN 228[5] tested by ISO 5164[6] is used as an example. The reproducibility of RON using this method is 0,7 RON at the EN 228 specification of 95,0 RON and the repeatability at the same level is 0,2 RON.
In this example, a supplier sells a batch of gasoline as compliant with EN 228 after certifying the product at their chosen laboratory as meeting specification. The supplier RON result is 95,1 against the specification of 95,0 and the batch is sold Free On Board (FOB). The sales contract specifies the representative sample as the shore tank sample which is stored appropriately and in sufficient volume for any follow-up testing.
The recipient purchases the batch and takes a sample to check for quality and the result from their chosen laboratory is determined as 94,7, which is off-specification but within 0,59R. Subclause 6.3.2 indicates that the product fails the specification limit with 95 % confidence only if it exceeds the specification by an amount greater than 0,59R; this is not the case for this example. It is common for the test against 0,59R to be the end of the procedure unless data from multiple batches is available.
In this case, despite the guidance in 6.3.2, the recipient is unhappy with their result and contacts the supplier regarding their result. Both are able to confirm and demonstrate that the results came from laboratories that have in-house SQC programs that are in control with respect to precision and bias for testing using ISO 5164 and that they regularly participate in industry PT schemes to confirm a lack of bias versus industry averages.
Under this scenario, 4.3.1 is applicable since both the supplier and the recipient results are available.
- Acceptability of results states that “when single results are obtained in two laboratories and their difference is less than or equal to R, the two results shall be considered as acceptable, and used to calculate the average . The average , rather than either single result separately, shall be used as the estimated value of the tested property”.
- First test the difference of the two values versus reproducibility to determine if the two results can be considered part of the same population of results.
|XS − XR| ≤ R = |95,1 − 94,7| ≤ R = 0,4 ≤ 0,7 so in this example both results are considered as valid. If this had not been the case then 4.3.1 gives the process to follow rejecting the suspicious results and how to gain acceptable results for comparison.
- After passing the comparison versus reproducibility, the average result is then calculated:
= (95,1 + 94,7)/2 = 94,9
- Now it is possible to calculate the 95 % confidence limits in which the true value µ lies using Formula (7), (8) or (9) as appropriate.
For a single lower limit, Formula (9) gives:
μ ≥ (− 0,42R), that is μ ≥ (94,9 − 0,294) or μ ≥ 94,6
- Based on the outcome from d), it can be concluded, with 95 % confidence, that the true value is no worse than 94,6. If the recipient still considers the product as in dispute, see Clause 7.
If only the recipient result is available, then 6.3.2 guides the comparison on conformance to specification and the confidence limit from Formula (9) in this example is simplified to μ ≥ (X − 0,59R).
If the recipient had a long-term relationship with the supplier and had data from multiple batches of product, 6.3.3 gives guidance on how to assess whether the confidence of product meeting specification is acceptable.
7.0 Dispute procedure
7.1 Resolve dispute by negotiation
Given the complexity of the procedures outlined in Clause 7 and Annex B, it is highly recommended that the supplier and recipient attempt to resolve the dispute through negotiation to arrive at mutually acceptable terms for settlement. Figure 3 is a flowchart summary of the steps outlined in Clause 7.
In order to engage in this procedure a pre-requisite is that there is sufficient amount of agreed-upon adjudication sample available.
7.1.1 Use of the test method or procedure in case of dispute
For a specification document that lists multiple test methods for the same property, or, a test method standard where multiple procedures for the same property are given, the designated referee method or procedure shall be given.
NOTE It is the responsibility of the standardization committee supervising either the specification or the test method standard to ensure a referee method or procedure for adjudication purposes is designated in the document.
In the event that a referee method or procedure is not specified, supplier and receiver shall agree on a specific method or procedure prior to execution of the procedures in this clause.
7.1.2 Dispute resolution procedure
If the supplier and the recipient cannot reach agreement regarding the specification conformance for the quality of the product on the basis of their existing results (see 6.3.3), then the dispute resolution procedures given in either this clause or Annex B shall be adopted. Note that in order to use Annex B, the degree of criticality, pc, is to be agreed upon in advance between supplier and recipient.
Each laboratory shall reject its original results and obtain at least three other acceptable results on their own check sample to ensure that the work has been carried out correctly under repeatability conditions. The average of the acceptable results in each laboratory shall then be computed, divergent results being discarded as indicated in 4.2.2. If the re-testing does not resolve the dispute, then continue as given below.
Let
| be the average of the supplier; | |
| be the average of recipient; | |
| A1 | be the upper limit of the specification; |
| A2 | be the lower limit of the specification. |
and shall be compared as follows with A1 and A2:
If ≤ A1 or ≥ A2
- product meets specification if (for R2, see 4.3.1);
- it cannot be stated with confidence whether the product does or does not comply with the specification limit if .
In the latter case, since it cannot be stated with confidence whether the product does or does not comply with the specification limit, resolution of the dispute may be achieved by negotiation.
If , then follow the dispute process
7.1.3 Dispute unresolved
If , and the dispute is unresolved, the two laboratories shall contact each other and compare their operating procedures and apparatus.
If the disagreement on product conformance status still remains and resolution of the dispute cannot be reached by negotiation, at least one additional third-party laboratory that is neutral, expert and accepted by the two parties in dispute shall be invited to assist in the dispute resolution. All parties (supplier, recipient, and third-party laboratory) shall first agree on a common adjudication sample. Subject to sample volume adequacy, each party shall obtain their own average to be used for adjudication purposes using at least three acceptable results for the adjudication sample in accordance with 4.2.2. If there is an insufficient amount of adjudication sample available, all parties shall agree to the number of test results that each party is to use to calculate their average to be used for adjudication purposes.
Suppose one additional third-party laboratory is used, and is the adjudication result from the third-party laboratory. If the difference between the most divergent laboratory adjudication result and the average of the two other laboratory adjudication results is less than or equal to R3 (see 4.3.1) the following procedure shall be adopted.
- If ≤ A1 or ≥ A2, product meets specification.
- If , product fails specification.
If more than one additional third-party laboratory is involved, and the difference between the most divergent laboratory adjudication result versus the average of all other laboratories' adjudication results is less than or equal to R3 (see 4.3.1), then the grand average from all adjudication results shall be used to settle the dispute as follows.
Let
XGRAND AVG be the grand average calculated from all adjudication results.
If XGRAND AVG ≤ A1 or ≥ A2, product meets specification, else product fails specification.
If the difference between the most divergent laboratory adjudication result and the average, , of the other adjudication results is more than R3, the following procedure shall be adopted.
- If ≤ A1 or ≥ A2, product meets specification.
- If , product fails specification.
Figure 3 — Flowchart for dispute procedure
7.1.4 Example of a dispute resolution
This subclause shows how to evaluate the procedure described so far in Clause 7. It is a continuation of the examples given in 6.3.4 using the same samples, results and customer/supplier. That example uses the RON specification within EN 228[5] tested by ISO 5164.[6] In this dispute procedure example, it is assumed that the use of 6.3.4 failed to bring the discussion between supplier and recipient to closure and that the contract did not specify a designated criticality level requiring the use of Annex B.
Having already demonstrated the ability of both laboratories to test, be in statistical control and a lack of persistent bias versus industry as part of 6.3.4, the supplier and recipient enter into the dispute procedure outlined in Clause 7.
All results taken to date as part of the process in 6.3.4 are rejected and assuming there is sufficient volume of the supplier and recipient check samples available, each then takes a minimum of 3 new results under repeatability conditions. Each pair of results taken are compared against method r.
The results for the supplier in their testing are 94,9 RON, 95,1 RON and 95,2 RON giving an average when rounded to the reporting precision plus 1 digit of 95,07 RON (this becomes the result).
The results for the recipient in their testing are 94,8 RON, 95,0 RON and 94,9 RON giving an average when rounded to the reporting precision plus one digit of 94,90 RON (this becomes the result).
The differences between the and averaged results shall meet the reduced reproducibility 0,84 × R2, where R2 is defined in Formula (21) below.
(21)
= 0,68 RON when rounded to an extra significant figure.
Clearly in this example the difference of 0,17 RON between supplier and recipient averages meets the 0,84 × R2 criteria and thus the results are accepted in the dispute process. Had this not been the case then the 3 paths forward are outlined in the dispute flow chart (see Figure 3).
To determine whether the product meets specification, the and averaged results are averaged and compared against specification. The average of 95,07 RON and 94,90 RON is 94,985 RON which is rounded to the reporting precision of 95,0 RON. This is compared to the specification of 95,0 RON thus passing the test.
In this case the dispute procedure outcome is that there is no dispute as the product meets specification (95,0 ON ≥ 95,0 ON).
Had this not been the case, and the receiver refuses to recognize that the product meets specification, then 7.4 outlines the path forward. This would involve agreeing a representative retained sample for additional testing, agreeing an external laboratory (E) to engage in the additional testing process alongside the supplier and recipient nominated laboratories and then repeating the multi testing process on the new sample. The divergence of averaged lab results is tested against a reduced reproducibility R3 defined in Formula (11).
(informative)
Explanation of formulae given in Clause 4
Let σ02 be the component of variance of results obtained under repeatability conditions.
Let σ12 be the component of variance due to interaction between laboratories and samples (errors which contribute to reproducibility errors). Then (σ02 + σ12) is the variance of results obtained under reproducibility conditions.
r is calculated as
and R is calculated as
where Z is the factor[4] for converting a standard deviation to a confidence limit, and which corresponds in this case to a double-sided 95 % probability level, having a value of 1,96. Table A.1 gives the critical values, Z, corresponding to a single sided probability, p, or to a double sided significance level 2 (1 − p).
Table A.1 — Critical values of the normal distribution
p | 0,70 | 0,80 | 0,90 | 0,95 | 0,975 | 0,99 | 0,995 |
Z | 0,524 | 0,842 | 1,282 | 1,645 | 1,960 | 2,326 | 2,576 |
2(1 − p) | 0,60 | 0,40 | 0,20 | 0,10 | 0,05 | 0,02 | 0,01 |
NOTE This table is reproduced from ISO 4259-1. |
Furthermore, the variance of means of k results obtained under repeatability conditions is σ02/k.
In a set of k such results, therefore, the variance of the difference between a single result and the mean of the remainder is:
(A.3)
and the 95 % confidence limit for the absolute difference is:
(A.4)
If the mean of k results is obtained in each of several laboratories, these laboratory means have a variance:
(A.5)
Let
(A.6)
The double-sided 95 % confidence limits for such means are:
Confidence limits for probability levels other than 95 % may be calculated by selecting the appropriate Z value from Table A.1 (with single- or double-sided probability as required) and multiplying by the conversion factor Z/1,96. For a single-sided probability of 95 %, Z = 1,64 and the conversion factor is 0,84.
In general, N laboratories obtain average results from k1, k2, ....kN results, respectively. The variance of the average of N such laboratory averages is:
(A.8)
Let
(A.9)
The double-sided 95 % confidence limits for such means then becomes:
Confidence limits for probability levels other than 95 %, single- or double-sided as required, can be calculated by selecting the appropriate value Z from Table A.1, and multiplying by the conversion factor Z/1,96.
In a set of N + 1 such averages, therefore, the variance of the difference between a single average of k results and the mean of the remaining N averages is:
(A.11)
The 95 % confidence limit, R3, for the absolute difference is therefore:
(A.12)
In the case of only two laboratory averages (when N = 1) this formula reduces to:
(A.13)
This annex presents an alternate procedure for dispute resolution between the supplier and recipient using a specified degree of specification criticality in view of the procedure in 7.3.
Some specifications, because of the product characteristic or the end use of the product, or both, require that the recipient has a high degree of assurance that the product meets or exceeds the quality level indicated by the specification level(s). For the purpose of this document, such specifications are called “critical specifications”.
Specifications that require assurance only that the quality is not substantially poorer than is indicated by the specification are called “non-critical specifications” for the purposes of this document.
For this document, the degree of criticality of a specification, denoted by pc, is defined to be the maximum probability (risk) that the recipient can tolerate of accepting a shipment which fails specifications. This pc is set as the maximum probability of accepting a shipment based on the limiting case where property in question is exactly at the specification limit. The risk borne by the supplier, that is, the recipient will reject a shipment which marginally meets or marginally does not meet the specification, will thus be 1 − pc. In order to use the procedure in this annex, pc is subject to prior agreement among the parties, and, like the specification limits and the test method, shall in that case be considered as an integral part of the specification.
This document considers specifications with a pc value of < 0,5 as critical specifications, while those with pc value ≥ 0,5 are considered to be non-critical specifications.
Execution of procedures in this annex assumes the parties in dispute (supplier and recipient) have agreed a priori to a pc as defined above.
While pc may be selected at any value that is agreed upon between supplier and recipient, this document recommends the maximum and minimum values for pc to be 0,95 (non-critical specification) and 0,05 (critical specification).
This step is described in 5.2.
This subclause provides general procedures to allow the parties in dispute (supplier and recipient) to accept or reject the product with regard to the specification when one or more acceptable results are available from one or more laboratories. Results are deemed acceptable if 6.1.2 to 6.1.4 are satisfied. If it is necessary for the recipient to take action after examining these results, the procedure specified in B.5 shall be adopted.
The procedures of B.5 assume a test method which follows the normal probability distribution, with no bias, and with repeatability, r, and reproducibility, R. They also assume that a degree of criticality, pc, has been agreed to in advance by supplier and recipient.
in the case of the upper limit of the specification A1,
in the case of a the lower limit of the specification A2,
and, in the case of a double limit (A1 and A2), both these conditions are satisfied.
The factor Z in these formulae is the value of the standard normal distribution corresponding to a probability, p (see Table A.1). Note that for critical specifications (pc < 0,5), Z has a negative value, and that the confidence 100 (1 − pc) % that the product meets the specification limit is greater than for non-critical specifications. The factor 0,361 is the reciprocal of [see Formula (A.2) used to convert reproducibility to a standard deviation].
If the reproducibility, R, is a function of the true value of the property in question, the value of R applied in Formula (B.1) is that which is appropriate for a true value of A1, whereas for Formula (B.2), R shall be computed assuming the true value is A2.
To engage in this procedure a pre-requisite is that there is a sufficient amount of agreed-upon adjudication sample available.
- Each laboratory shall reject its original results and obtain at least three other acceptable results on the adjudication sample to ensure that the work has been carried out under repeatability conditions. Result(s) from each laboratory is considered to be acceptable if 6.1.2 to 6.1.4 are met, and, in the case of more than one result, if 4.2.2 is met. The average of the acceptable results in each laboratory shall then be computed, divergent results being discarded as indicated in 4.2.2. The supplier's and recipient's averages will be denoted and , respectively.
- If and are acceptable in terms of reproducibility (see 4.3.1), and if satisfies Formula (B.1) or Formula (B.2) or both, as appropriate, with R replaced by R2 of 4.3.1 if necessary, then the product meets the specification.
- If and are acceptable in terms of reproducibility (see 4.3.1) and if fails either Formula (B.1) or Formula (B.2), as appropriate, with R replaced by R2 of 4.3.1 if necessary, then the product fails the specification.
- In the event that the difference in laboratory means, | − |, exceeds R (for a single result from each laboratory), or R2 of 4.3.1 (for multiple results from each lab), and if the dispute cannot otherwise be settled at this point, B.5.7 shall be applied.
- In the case of unacceptable laboratory averages, the two laboratories shall contact each other and compare their operating procedures and apparatus. Following these investigations, a correlation test between the two laboratories shall be carried out on the two check samples. The average of at least three acceptable results shall be computed, in each laboratory, and these averages compared as given in B.5.3 to B.5.6
- If the disagreement remains, a third laboratory (neutral, expert and accepted by the two parties) shall be invited to carry out the test using a third sample. Suppose is the average of the three or more acceptable results of the third laboratory. If the averages, and are acceptable in terms of reproducibility (see 4.3.1) then:
- if satisfies Formula (B.1) or Formula (B.2) or both, as appropriate, with R replaced by , then the product meets the specification [see Formula (10) with N = 3];
- if fails either Formula (B.1) or Formula (B.2), as appropriate, with R replaced by , then the product fails the specification.
- If the averages, and are not acceptable in terms of reproducibility (see 4.3.1), then the most divergent laboratory average shall be rejected, and the average of the two remaining averages are denoted as . R4 shall be recomputed based on the number of test results obtained by the remaining two laboratories, and becomes identical with R2 [Formula (10)].
- If satisfies Formula (B.1) or Formula (B.2) or both, as appropriate, with R replaced by then the product meets the specification.
- If fails either Formula (B.1) or Formula (B.2), as appropriate, with R replaced by , then the product fails the specification.
(informative)
General approach to bias assessment using multiple materials
This annex describes the general approach to determine if a bias correction can statistically improve the agreement between results obtained from two different test methods that purport to measure the same property over multiple materials.
Two test methods that purport or claim to measure the same property but are based on technically different analytical principles will, in general, yield results that are statistically different in the long run over the intersecting scope and range of materials for the methods.
A common approach practised by users is to use simple (or, ordinary) linear regression to develop correlation between the two methods, without giving due consideration to the precision of each test method. This approach is statistically incorrect since ordinary linear regression assumes that the values used for the independent variable are known error free and the precision associated with the dependent variable are constant across the region of regression. Neither of the assumptions is realistic for test method correlation.
A statistically correct and pragmatic approach is to ask the question: “can a general linear mathematical correction (model) be applied to results from one of the test methods such that a statistically observable improvement in the agreement between the two methods can be achieved?”. The rationale for limiting this correction to a linear model is based on the argument that if a more complex model form is required, the two methods are likely not measuring the same property directly.
The general approach to answer this question is outlined below.
- Define a sample set comprising the material types and range for the property of interest.
- Conduct an ILS for each test method using this sample set.
- Perform a regression analysis with a technique known as ReXY (Regression with errors in X and Y) using the outlier-free data from each test method and the respective standard errors for each material in the ILS. These standard errors can be calculated using the published R of each test method, but only on the condition that R derived from the ILS is not statistically different from the published R.
- Test the slope and intercept from c) for statistical significance and only include the term(s) that is significant for the bias correction model.
- Statistically compare the agreement between the uncorrected test results versus the corrected results using the model from d), which, can be no correction, a constant correction, or a constant + proportional correction to be made. There is more than one approach to perform the statistical comparison. The simplest is to compute the sum of squares of the uncorrected differences and compare this to the sum of squares of corrected differences using an F-test for significance.
Additional checks such as the adequacy of the property variation in the sample set relative to the test method precision, the adequacy of correlation, and the distribution of corrected differences should also be part of this study.
A detailed discussion on the subject matter is beyond the intended scope of this informative annex. Interested readers are referred to ISO 4259-5[3] for an in-depth coverage of the statistical methodology.
See ISO 4259-1:2024, Annex I, for a glossary of the variables and constants used in this document and ISO 4259-1.
[1] ASTM D6300, Standard Practice for Determination of Precision and Bias Data for Use in Test Methods for Petroleum Products and Lubricants
[2] ASTM D3244, Standard Practice for Utilization of Test Data to Determine Conformance with Specifications
[3] ISO 4259-5, Petroleum and related products — Precision of measurement methods and results — Part 5: Statistical assessment of agreement between two different measurement methods that claim to measure the same property
[4] Smith I.J. J. Inst. Pet. 1963, 49 pp. 155–162
[5] EN 228, Automotive fuels — Unleaded petrol — Requirements and test methods
[6] ISO 5164, Petroleum products — Determination of knock characteristics of motor fuels — Research method