prEN ISO 18227: Environmental solid matrices - Determination of elemental composition by X-ray fluorescence spectrometry (ISO/DIS 18227:2024)

ISO/DIS 18227:2024(en)

ISO/TC 190/SC 3/WG 1

Date: 2024-07-29

Secretariat: DIN

Environmental solid matrices — Determination of elemental composition by X-ray fluorescence spectrometry

© ISO 2024

All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of the requester.

ISO copyright office

CP 401 • Ch. de Blandonnet 8

CH-1214 Vernier, Geneva

Phone: +41 22 749 01 11

Email: copyright@iso.org

Website: www.iso.org

Published in Switzerland

Contents

Foreword v

Introduction vi

1 Scope 1

2 Normative references 1

3 Terms and definitions 1

4 Safety remarks 3

5 Principle 3

6 Apparatus 3

7 Reagents 4

8 Interferences and sources of error 5

9 Sample preparation 5

9.1 General 5

9.2 Drying and determination of dry mass 6

9.3 Preparation of pressed pellet 6

9.4 Preparation of fused beads 6

10 Procedure 7

10.1 Analytical measurement conditions 7

10.1.1 Wavelength dispersive instruments 7

10.1.2 Intensities and background corrections 7

10.1.3 Counting time 7

10.1.4 Energy dispersive instruments 7

10.1.5 Intensities and background corrections 7

10.2 Calibration 7

10.2.1 General 7

10.2.2 General calibration procedure 8

10.2.3 Internal standard correction using Compton (incoherent) scattering method 8

10.2.4 Fundamental parameter approach 9

10.2.5 Fundamental or theoretical influence coefficient method 9

10.2.6 Empirical alpha correction 10

10.2.7 Calibration procedure for trace elements using the pressed pellet method 10

10.2.8 Calibration procedure for major and minor oxides using the fused bead method 12

10.3 Analysis of the samples 13

11 Quality control 13

11.1 Drift correction procedure 13

11.2 Blank test 14

11.3 Reference materials 14

12 Calculation of the result 14

13 Test report 14

Annex A (informative) Semi-quantitative screening analysis of waste, sludge and soil samples 15

A.1 Principle 15

A.2 Energy dispersive (ED) or wavelength dispersive (WD) X-ray fluorescence spectrometer 15

A.3 Sample preparation 15

A.4 Procedure 15

A.4.1 Analytical measurement conditions and calibration 15

A.4.2 Validation 16

A.4.3 Analysis 16

A.5 Calculation and evaluation 17

A.6 Quality control 17

A.7 Safety remarks 17

A.8 Test report 17

Annex B (informative) Examples for operational steps of the sample preparation for soil and waste samples 18

B.1 General 18

B.2 Preparation of liquid samples 18

B.3 Preparation of powder samples 18

B.4 Impregnation on powders 18

B.4.1 Impregnation on aluminium oxide 18

B.4.2 Impregnation on activated carbon 18

Annex C (informative) Suggested analytical lines, crystals and operating conditions 23

Annex D (informative) List of reference materials applicable for XRF analysis 25

Annex E (informative) Validation 26

E.1 Result of validation 26

E.2 Discussion of the results 37

Bibliography 38

Foreword

ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.

The procedures used to develop this document and those intended for its further maintenance are described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).

ISO draws attention to the possibility that the implementation of this document may involve the use of (a) patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent rights in respect thereof. As of the date of publication of this document, ISO [had/had not] received notice of (a) patent(s) which may be required to implement this document. However, implementers are cautioned that this may not represent the latest information, which may be obtained from the patent database available at www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.

Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions related to conformity assessment, as well as information about ISO's adherence to the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.

This document was prepared by Technical Committee ISO/TC 190, Soil quality, Subcommittee SC 3, Chemical and physical characterization.

This second edition cancels and replaces the first edition (ISO 18227:2014), which has been technically revised.

The main changes compared to the previous edition are as follows:

— the contents of the two almost identical standards ISO 18277:2014 and EN 15309:2007 have been combined;

— normative references have been revised.

Any feedback or questions on this document should be directed to the user’s national standards body. A complete listing of these bodies can be found at www.iso.org/members.html.

Introduction

X-ray fluorescence spectrometry is a fast and reliable method for the quantitative analysis of the total content of certain elements within different matrices.

The quality of the results obtained depends very closely on the type of instrument used, e.g. bench top or high performance, energy dispersive or wavelength dispersive instruments. When selecting a specific instrument several factors have to be considered, such as the matrices to be analysed, elements to be determined, detection limits required and the measuring time. The quality of the results depends on the element to be determined and on the surrounding matrix.

Due to the wide range of matrix compositions and the lack of suitable reference materials in the case of inhomogeneous matrices such as waste, it is generally difficult to set up a calibration with matrix- matched reference materials.

Therefore this standard describes two different procedures:

— a quantitative analytical procedure for homogeneous solid waste, soil and soil-like material in the normative part. The calibration is based on matrix-matched standards;

— an XRF screening method for solid and liquid material as waste, sludge and soil in Annex A which provides a total element characterization at a semi-quantitative level. The calibration is based on matrix-independent calibration curves, previously set up by the manufacturer.

Environmental solid matrices — Determination of elemental composition by X-ray fluorescence spectrometry

1.0 Scope

This document specifies the procedure for a quantitative determination of major and trace element concentrations in homogeneous solid waste, soil, soil-like material and sludge by energy dispersive X-ray fluorescence (EDXRF) spectrometry or wavelength dispersive X-ray fluorescence (WDXRF) spectrometry using a calibration with matrix-matched standards.

This document is applicable for the following elements: Na, Mg, Al, Si, P, S, Cl, K, Ca, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, As, Se, Br, Rb, Sr, Y, Zr, Nb, Mo, Ag, Cd, Sn, Sb, Te, I, Cs, Ba, Ta, W, Hg, Tl, Pb, Bi, Th and U. Concentration levels between approximately 0,000 1 % and 100 % can be determined depending on the element and the instrument used.

2.0 Normative references

There are no normative references in this document.

3.0 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminology databases for use in standardization at the following addresses:

— ISO Online browsing platform: available at https://www.iso.org/obp

— IEC Electropedia: available at https://www.electropedia.org/

3.1

absorption edge

jump of the mass absorption coefficient at a specific wavelength or energy

3.2

absorption of X-rays

loss of intensity of X-rays by an isotropic and homogenous material as described by the Bouger-Lambert law

3.3

analytical line

specific characteristic X-ray spectral line of the atom or ion of the analyte used for determination of the analyte content

3.4

continuous radiation

electromagnetic radiation produced by the acceleration of a charged particle, such as an electron, when deflected by another charged particle, such as an atomic nucleus

3.5

Compton-line

spectral line due to incoherent scattering (Compton-effect) occurring when the incident X-ray photon strike an atom without promoting fluorescence

Note 1 to entry: Energy is lost in the collision and therefore the resulting scattered X-ray photon is of lower energy than the incident X-ray photon.

3.6

drift correction monitor

physically stable sample used to correct for instrumental drift

3.7

emitted sample X-rays

radiation emitted by sample consisting of X-ray fluorescence radiation and scattered primary X-rays

3.8

fused bead

analyte sample prepared by dissolution in a flux

3.9

liquid sample

analyte sample submitted as a solution for direct measurement in the sample cup

3.10

mass absorption coefficient

constant describing the fractional decrease in the intensity of a beam of X-radiation as it passes through an absorbing medium

Note 1 to entry: This is expressed in units of cm2/g.

Note 2 to entry: The mass absorption coefficient is a function of the wavelength of the absorbed radiation and the atomic number of the absorbing element.

3.11

polarized excitation X-ray spectrometer

energy dispersive X-ray spectrometer where the excitation is performed by polarized radiation and the emitted X-ray fluorescence radiation is detected along the direction of polarization

3.12

powder sample

analyte sample submitted as a powder for direct measurement in the sample cup

3.13

precision

closeness of agreement of results obtained by applying the method several times under prescribed conditions

[SOURCE: ISO 5725‑2:2019, x.xx]

3.14

pressed pellet

analyte sample prepared by pressing milled material into a disk

3.15

primary X-ray

X-ray by which the sample is radiated

3.16

quality control sample

stable sample with known contents, e.g. certified reference material (CRM) used to monitor instrument and calibration performance

3.17

X-ray fluorescence radiation

emission of characteristic X-rays from a sample that has been bombarded by high-energy X-rays or gamma rays

4.0 Safety remarks

Anyone dealing with waste and sludge analysis has to be aware of the typical risks that this kind of material presents irrespective of the parameter to be determined. Waste and sludge samples can contain hazardous e.g. toxic, reactive, flammable, and infectious substances, which could potentially undergo biological and/or chemical reaction. Consequently, it is recommended that these samples should be handled with special care. The gases that can be produced by microbiological or chemical activity are potentially flammable and pressurize sealed bottles. Bursting bottles are likely to result in hazardous shrapnel, dust and/or aerosol. National regulations should be followed with respect to all hazards associated with this method.

The X-ray fluorescence spectrometer shall comply with European/international and national regulations relevant to radiation protection.

The person responsible for managing or supervising the operation of X-ray equipment shall provide evidence of his knowledge of radiation protection according to national regulations.

5.0 Principle

After a suitable preparation, if necessary, the sample is introduced into an XRF-spectrometer and excited by primary X-rays. The intensities of the secondary fluorescent energy lines specific for each element are measured and the elemental composition of the sample is determined by reference to previously established calibration graphs or equations and applying corrections for inter-element effects. The calibration equations and inter-element corrections are established using pure reagents and/or series of internal or reference materials providing they meet all the requirements of the relevant preparation technique.

6.0 Apparatus

6.1 X-ray fluorescence spectrometer, shall be able to analyse the elements according to the scope of this document.

The following types of X-ray fluorescence spectrometers are applicable:

— energy dispersive X-ray fluorescence (EDXRF) spectrometer that achieves the dispersion of the emitted X-ray fluorescence radiation by an energy dispersive detector;

— wavelength dispersive X-ray fluorescence (WDXRF) spectrometer that achieves the dispersion of the emitted X-ray fluorescence radiation by diffraction by a crystal or a synthetic multilayer.

The spectrometer consists of a number of components:

— primary X-ray source, an X-ray tube with a high voltage generator;

— a sample holder;

— detector unit including electronic equipment;

— source modifiers to modify the shape or intensity of the source spectrum or the beam shape (such as source filters, secondary targets, polarizing targets, collimators, focussing optics, etc.).

The detector unit is different for WDXRF and for EDXRF spectrometers. WDXRF spectrometers take advantage of the dispersion of the emitted radiation by scattering by a crystal or a synthetic multilayer. The detector does not need to be capable of energy discrimination. EDXRF spectrometers use an energy dispersive detector. Pulses of current from the detector, which are a measure of the energy of the incoming X-rays, are segregated into channels according to energy using a multi-channel analyser (MCA). The spectrometer is capable to measure under vacuum, helium-atmosphere (7.3) or nitrogen- atmosphere (7.4).

NOTE 1 The use of a high-energy X-ray tube increases the potential for losses of volatile analytes from samples by heating in the spectrometer during analysis.

NOTE 2 The new generation of EDXRF spectrometers takes advantage of the polarizing target theory resulting in a significant decrease of the background scattering, and therefore lower limits of detection can be achieved (comparable to WDXRF).

6.2 Analytical balance, readable and accurate to 0,001 g.

6.3 Drying oven, thermostatically controlled and capable of maintaining a temperature of (105 ± 5) °C.

6.4 Grinding mill, capable of grinding dried materials to a required particle size without contaminating the samples with compounds to be determined, preferable with walls made of agate, corundum or zircon.

6.5 Pellet preparation equipment, manual or automatic pellet press, capable of providing a pressure of at least 100 kN.

6.6 Aluminium cup: supporting backing cup for pressed pellets.

6.7 Fusion apparatus: electric, gas or high frequency induction furnace that can be heated up to a fixed temperature of between 1 000 °C and 1 250 °C.

6.8 Fusion crucibles: crucibles made of non-wetting platinum alloy (Pt 95 %; Au 5 % is suitable).

Lids, if used, shall be made from platinum alloy.

NOTE Certain metal sulphides (so called platinum poisons) affect the platinum crucibles in which the sample is melted.

6.9 Casting moulds: non-wetting platinum alloy (Pt 95 %; Au 5 % is suitable).

7.0 Reagents

The reagents mentioned are used as carrier material.

7.1 Binder: liquid or solid binder free of analytes of interest.

Solid materials can contain a certain amount of moisture, which shall be compensated for.

NOTE Different type of binders may be used. A binder commonly used is wax.

7.2 Flux: solid flux free of analytes of interest.

Solid materials can contain a certain amount of moisture, which shall be compensated for (e.g. ISO 12677 for compensation for moisture in flux).

NOTE Different type of fluxes may be used. Fluxes commonly used are lithium metaborate, lithium tetraborate or mixtures of both.

7.3 Helium, purity ≥ 99,996 %.

7.4 Nitrogen, purity ≥ 99,996 %.

8.0 Interferences and sources of error

The container in which the sample is delivered and stored can be a source of error. Its material shall be chosen according to the elements to be determined.

NOTE Elemental Hg can penetrate polyethylene walls very rapidly in both directions. In the case of glass containers, contamination can be observed for some elements e.g. Al, As, Ba, Ce, K, Na and Pb.

Interferences in X-ray fluorescence spectrometry are due to spectral line overlaps, matrix effects, spectral artefacts and particle size or mineralogical effects.

Spectral line overlaps occur when an analytical line cannot be resolved from the line of a different element. Corrections for these interferences are made using the algorithms provided with the software.

Matrix effects occur when the X-ray fluorescence radiation from the analyte element is absorbed or enhanced by other elements in the sample before it reaches the detector. In the case of complex matrices these effects generally have to be corrected.

Spectral artefacts e.g. escape peaks, sum peaks, pulse pile up lines, dead time, and continuous radiation correction, are accounted for by the provided software. Spectral artefacts differ for energy dispersive and wavelength dispersive XRF spectrometry.

Particle size effects can be reduced by milling the sample, and both particle size and mineralogical effects can be eliminated by preparing bead samples. It is vital for quantitative analysis that the same sample preparation procedure is applied to both the standards and the samples to be analysed.

9.0 Sample preparation

9.1 General

In analysis by XRF spectrometry the sample preparation step is crucial as the quality of the sample preparation strongly influences the accuracy of the results.

For quantitative analysis of solid samples, pressed pellets or fused beads have to be prepared. The application of the pressed pellet method is recommended for the quantification of trace elements and mandatory for the quantification of volatile elements, and the fused bead method for the determination of non-volatile major and minor elements.

NOTE 1 The preparation of fused beads eliminates effects due to particle size and mineralogy.

The conditions of the preparation of fused beads shall be adapted to the matrix properties. Otherwise the preparation of fused beads can be difficult or can cause problems in case of waste-like matrices such as sludges.

For a given calibration the same preparation method shall be used throughout, for both samples and standards.

NOTE 2 Depending on the sample type other sample preparation methods can be applied according to Annex B.

For precise quantitative measurements, homogeneous and representative test portions are necessary. If not otherwise specified, pre-treatment and preparation of test portions should be carried out according to the appropriate clauses of e.g. ISO 11464, EN 15002 or EN 16179. The particle size of the sample can strongly affect the precision of the measurement. The particle size should preferably be smaller than 150 µm.

NOTE 3 Particle size smaller than 80 µm is recommended for the analysis of low atomic mass elements when using the pressed pellet method.

9.1.1 Drying and determination of dry mass

If not otherwise specified, the determination of the dry mass should be carried out according to e.g. ISO 11465 or EN 15934.

9.1.2 Preparation of pressed pellet

After drying (6.3) and milling or grinding the sample (6.4), a pellet is prepared in the pellet press (6.5). Before pressing, the sample shall be mixed and homogenized with a binder (7.1) in a ratio of sampler:binder of 10:1 by weight. For the preparation of 40 mm in diameter pellets, about 10,0 g of sample is taken; for 32 mm in diameter pellets about 4,5 g of sample is required. The amount of binder in the pellet shall be taken into account for the dilution factor. It is recommended to press the sample in an aluminium cup (6.6) as support.

NOTE 1 Different types of binders can be used. A binder commonly used is wax. In the case of a liquid binder the pellet is placed in an oven to evaporate organic solvent.

NOTE 2 Different dilution factors can be used.

9.1.3 Preparation of fused beads

After drying (6.3) and milling or grinding the sample (6.4), a fused bead is prepared using the fusion apparatus (6.7).

Ignite the sample at 1 025 °C ± 25 °C until constant mass is reached. Determine the loss on ignition at the chosen temperature to correct for volatile elements and/or compounds being released during ignition of the sample.

NOTE 1 The ignition temperature can vary depending on the sample matrix.

Because of the wide applicability of the fused bead technique, various fluxes and modes of calibration are permitted providing they have been demonstrated to be able to meet certain criteria of reproducibility, sensitivity and accuracy.

For application of alkaline fusion technique (e.g. selection of flux, fusion temperature, and additives) e.g. ISO 14869‑2 or CEN/TR 15018 should be used.

NOTE 2 Fluxes commonly used are lithium metaborate, lithium tetraborate or mixtures of both.

NOTE 3 Loss of volatile elements, e.g. As, Br, Cd, Cl, Hg, I, S, Sb, Se, and Tl can occur during the fusion process. Also, Cu can be volatile if a bromide-releasing agent is used.

The flux (7.2) is added to the ignited material in a dilution ratio of sample:flux of 1:5 by weight. For the preparation of 40 mm in diameter beads, about 1,6 g of ignited sample is taken; for 32 mm in diameter beads about 0,8 g of ignited sample is required. The amount of flux in the bead shall be taken into account for the dilution factor. The same sample preparation procedure and ratio of sample to flux shall be used for samples and standards. The beads produced should be visually homogeneous and transparent.

NOTE 4 Non-ignited material can be used to prepare beads but, nevertheless, loss of ignition needs to be determined and needs to be taken into account in the calculation of the results. It should be noted that non-ignited material can contain compounds that can damage the platinum crucibles during fusion.

NOTE 5 Different dilution factors can be used.

After fusion in a platinum-gold crucible (6.8) the melt is poured into a casting mould (6.9) to make a bead.

Beads can deteriorate because of adverse temperature and humidity conditions, so it is recommended that beads are stored in desiccators.

10.0 Procedure

10.1 Analytical measurement conditions

10.1.1 Wavelength dispersive instruments

The analytical lines to be used and the suggested operating conditions are given in Table C.1. The settings are strongly dependent on the spectrometer configuration, e.g. the type of X-ray tube (Rh, Cr), tube power, available crystals, and type of collimators.

10.1.2 Intensities and background corrections

For the determination of trace elements the measured intensities have to be background-corrected. The measured background positions should be free of spectral line interferences. The net peak intensity I, expressed as the number of counts per second of the element of interest, is calculated as the difference between the measured peak intensity of the element and the background intensity:

(1)

where

 

is the count rate of the element i, expressed as the number of counts per second;

 

is the background count rate of the element i, expressed as the number of counts per second.

10.1.3 Counting time

The minimum counting time is the time necessary to achieve an uncertainty (2 %), which is less than the desired precision of the measurement. Choose a reference material with a concentration level in the middle of the working range and measure the count rate. The counting time for each element can be calculated according to Formula (2):

(2)

where

 

is the total counting time for the peaks and background, in seconds; ;

 

is the relative target precision at a confidence level of 95 %, expressed as percentage.

10.1.4 Energy dispersive instruments

The analytical lines to be used and the suggested operating conditions are given in Table C.2. The settings are strongly dependent on the spectrometer configuration, e.g. type of X-ray tube (Rh, Pd), tube power, available targets, and type of filters.

10.1.5 Intensities and background corrections

Deconvolution of the spectra and background correction are needed when analysing the samples with overlapping lines. Usually, XRF instruments are supplied with a specific software module for that purpose.

10.2 Calibration

10.2.1 General

The calibration procedure is similar for energy dispersive and wavelength dispersive techniques. In general calibration is established by using matrix-adapted reference materials. The calibration equations and inter-element corrections are calculated by the software of the instrument. An accuracy check is performed with CRMs or samples with known composition.

Different procedures for correcting matrix effects can be used according to the analytical accuracy required:

— the scattered radiation method is based on the principle that the intensities of the analyte line and of the Compton line are affected in the same proportion due to the overall mass absorption coefficient of the sample. This linear relationship holds when all analytes are at low concentrations (trace elements) and their absorption coefficients are not affected by an adjacent absorption edge. In this case an internal Compton correction can be used. Aside from that, a correction method using the Compton intensity with mass absorption coefficients (MAC) is also applicable. In this method, the intensities of the major elements are measured to apply a jump edge correction for the analysed trace elements;

— correction using the fundamental parameter approach;

— correction using theoretical correction coefficients (alphas) taking basic physical principles, instrumental geometry etc. into account;

— correction using empirical correction coefficients (alphas) based on regression analysis of standards with known elemental concentrations.

10.2.2 General calibration procedure

For calibration purposes the measurement of analyte lines of samples of known composition is needed. Formula (3) implies a linear relationship between the intensity and the concentration.

(3)

where

 

is the concentration of the element of interest, expressed as mg/kg or percentage dry matter;

 

is the offset of the calibration curve;

 

is the slope of the calibration curve;

 

is the net intensity of the element of interest, expressed as counts per second.

Matrix effects have to be taken into account in X-ray spectrometry according to Formula (4):

(4)

where, M is the correction term due to the matrix effects.

The matrix effect correction term can consist of an internal standard Compton correction method or can be calculated from mathematical models.

10.2.3 Internal standard correction using Compton (incoherent) scattering method

The measured intensity of incoherent scattering can be used directly to compensate for matrix effects or indirectly for the determination of the effective mass absorption coefficient to correct for matrix effects. The compensation for matrix effects is based on a combination of sample preparation and experimental intensity data but not on fundamental and experimental parameters.

The Compton scatter method can be expressed as:

(5)

where

 

is the concentration of the element of interest i of the sample, expressed as mg/kg or percentage dry matter;

 

is the concentration of the element of interest i of the calibration reference material, expressed as mg/kg or percentage dry matter;

 

is the intensity of the incoherent Compton-line element of the calibration reference material, expressed as counts per second;

 

is the intensity of the element of interest i of the calibration reference material, expressed as counts per second;

 

is the intensity of the element of interest i of the sample, expressed as counts per second;

 

is the intensity of the incoherent Compton-line of the sample, expressed as counts per second.

10.2.4 Fundamental parameter approach

The fundamental parameter approach uses the physical processes forming the basis of X-ray fluorescence emission and scattering to construct a theoretical model for the correction of matrix effects in practice. The correction term M is calculated from first principle expressions. These are derived from basic X-ray physics and contain physical constants and parameters that include absorption and scattering coefficients, fluorescence yield, primary spectral distributions and spectrometry geometry. The use of scattered radiation (Compton and/or Rayleigh) allows the determination of matrix effects caused by sample elements that cannot be measured directly. The calculation of analyte concentrations in samples is based on making successively better estimates of composition by an iteration procedure. These iteration cycles are performed until the difference between the compared results is below a defined value.

NOTE The algorithm used for the procedure is usually implemented in the manufacturer’s software.

10.2.5 Fundamental or theoretical influence coefficient method

The fundamental influence coefficient method encompasses any mathematical expression relating emitted intensities and concentrations in which the influence coefficients are defined and derived explicitly in terms of fundamental parameters.

The calculation of the concentration from the intensities is performed by linear regression whereby the net intensities are corrected for the present matrix effects. For each element the concentration is calculated according to Formulae (6) and (7):

(6)

(7)

where

 

is the concentration of the element of interest i of the sample, expressed as mg/kg or percentage dry matter;

 

is the concentration of the element of interest i of the calibration reference material, expressed as mg/kg or percentage dry matter;

 

is the intensity of the incoherent Compton-line of the sample, expressed as counts per second;

 

is the intensity of the element of interest i of the sample, expressed as counts per second;

 

is the concentration of the matrix element j of the calibration reference material, expressed as mg/kg or percentage dry matter;

 

is the concentration of the matrix element j of the sample, expressed as mg/kg or percentage dry matter;

 

M

is the matrix correction term;

 

is the correction coefficient (called alphas) calculated from theory, although some approximations are involved.

Different types of alpha coefficient exist, but all of them are calculated without reference to experimental data; they are calculated using intensity data resulting from a fundamental parameter expression. The alpha coefficients vary as a function of sample composition and are calculated by an iterative process.

10.2.6 Empirical alpha correction

Empirical alphas are obtained experimentally using the regression analysis of data from reference materials in which the elements to be measured are known and the total concentration range is covered. Best results are achieved when the samples and reference materials are of similar composition. Thus, empirical alphas are based strictly on experimental data and do not take fundamental and instrumental parameters into account. Different models can be applied, but generally they are based on Formulae 6 and 7, where the correction term for matrix effects is a function of concentrations.

The empirical alphas are only applicable for a limited concentration range and a well-defined analytical method where the matrices of samples and standards are similar. The reference materials used should contain each analyte together with fairly wide concentration ranges of each matrix element. Poor analytical results are obtained when inappropriate combinations of analytes are chosen. A large number of reference materials have to be analysed to define the alphas (rule of thumb: minimum of 3 times the number of parameters to be calculated).

10.2.7 Calibration procedure for trace elements using the pressed pellet method

The pressed pellet method is used to determine the concentrations of trace elements.

Select calibration standards with a similar composition as the samples under investigation containing the elements of interest and covering the concentration range of interest. The use of reference materials from different recognized producers is recommended (see Annex D) or synthetic mixtures of oxides can be prepared. The element concentrations shall vary independently in the standards. If the calibration covers many elements in a wide range of concentrations, a large number of calibration samples can be necessary.

Prepare pressed pellets from the selected calibration standards according to 9.3.

Define the analytical measurement method for EDXRF or WDXRF as described in 10.1.

Start up the XRF equipment according to the instrument manufacturer’s manual and measure the calibration standards using the defined measurement method. All measurements shall be performed under vacuum or helium-atmosphere or nitrogen-atmosphere.

NOTE It is important to note that the pressed pellet method is not ideal for the determination of major elements, but these elements are measured so that alpha corrections can be applied to some elements of interest.

Follow the guidelines in the instrument manufacturer’s manual to perform the regression, the background correction, the line overlap correction and the matrix corrections for all elements under consideration.

In Table 1 the possible spectral line overlaps are indicated (dependent on the configuration of the instrument) and also the matrix correction method that can be applied. For trace elements with an absorption edge above the absorption edge of iron, a Compton internal standard correction can be applied. Otherwise a theoretical alpha correction or correction for the absorption edge should be performed (for these corrections all elements in the sample have to be analysed).

Depending on the type of instrument and the software programs available, alternative correction methods can be applied. Validation of the final calibration curves shall demonstrate the accuracy of the method.

Perform the regression calculation and verify that the correlation factors are within the limits of accuracy required.

Table 1 — Suggested analytical lines, spectral line overlaps and correction methods

Element

Line

Spectral line overlap

Type of matrix correction method

Na

ZnLβ

Alpha or FP

Mg

AsLα

Alpha or FP

Al

BrLα

Alpha or FP

Si

Alpha or FP

P

Alpha or FP

S

CoKα PbMα NbLβ

Alpha or FP or MAC

Cl

Alpha or FP or MAC

K

Alpha or FP

Ca

Alpha or FP

Ti

BaLα ΙLβ

Alpha or FP

V

Ti Kβ

Alpha or FP or MAC

Cr

VKβ PbLα

Alpha or FP or MAC

Mn

CrKβ

Alpha or FP

Fe

MnKβ

Alpha or FP

Co

FeKβ

Alpha or FP or MAC

Ni

CoKβ

Compton or FP or MAC

Cu

TaLα ThLβ

Compton or FP or MAC

Zn

WLα

Compton or FP or MAC

As

Kα Kβ

PbLα BrKα

Compton or FP or MAC

Se

Compton or FP or MAC

Br

AsKβ

Compton or FP or MAC

Rb

ULα BrKβ

Compton or FP or MAC

Sr

ULα

Compton or FP or MAC

Y

RbKβ

Compton or FP or MAC

Zr

SrKβ

Compton or FP or MAC

Nb

YKβ ULβ

Compton or FP or MAC

Mo

ZrKβ ULβ

Compton or FP or MAC

Ag

Kα Lα

CrKβ

Compton or FP or MAC Alpha or FP

Cd

AgLβ

Compton or FP or MAC Alpha or FP

Sn

Kα Lα

CoKα

Compton or FP or MAC Alpha or FP or MAC

Sb

Kα Lβ

CoKβ

Compton or FP or MAC Alpha or FP or MAC

Te

Kα Lα

SnLβ

Compton or FP or MAC Alpha or FP or MAC

I

Compton or FP or MAC Alpha or FP or MAC

Cs

Kα Lα

ZnKα ILβ

Compton or FP or MAC Alpha or FP or MAC

Ba

Kα Lα

TiKα ILβ CuKβ

Compton or FP or MAC Alpha or FP or MAC

Ta

CuKα NiKβ

Compton or FP or MAC

W

TaLn

Compton or FP or MAC

Hg

WLβ

Compton or FP or MAC

Tl

PbLβ

Compton or FP or MAC

Pb

ThLα BiLβ SnKα

Compton or FP or MAC

Bi

TaLγ

Compton or FP or MAC

Th

BiLβ PbLβ

Compton or FP or MAC

U

BrKβ RbKα

Compton or FP or MAC

10.2.8 Calibration procedure for major and minor oxides using the fused bead method

The fused bead method is used to determine the concentrations of major and minor elements.

Select calibration standards with a similar composition as the samples under investigation containing the elements of interest and covering the total concentration range of interest. The use of reference materials from different recognized producers is recommended (see Annex D) or synthetic mixtures of oxides can be prepared. The element concentrations shall vary independently in the samples. If the calibration covers many elements in a wide range of concentrations, a large number of calibration samples can be necessary.

Prepare fused beads from the selected calibration standards according to 9.4.

NOTE Due to a higher dilution factor for fused beads the limit of detection of the different elements is higher than those for pressed pellets.

Define the analytical measurement method for EDXRF or WDXRF as described in 10.1.

Start up the XRF equipment according to the instrument manufacturer’s manual and measure the calibration standards using the defined measurement method. All measurements shall be performed under vacuum or helium-atmosphere or nitrogen-atmosphere.

In the calibration program all the elements of the reference materials have to be defined as oxides. The concentrations can be reported as oxides or as element concentrations.

Follow the guidelines in the instrument manufacturer’s manual how to perform the regression, the background correction, the line overlap correction and the matrix corrections for all elements under consideration.

In Table 1 the possible spectral line overlaps are indicated (dependent on the configuration of the instrument). For all elements an alpha correction method using theoretical alphas should be applied.

Depending on the type of instrument and the software programs available, alternative correction methods can be applied. Validation of the final calibration curves shall demonstrate the accuracy of the method.

Perform the regression calculation and verify that the correlation factors are within the limits of accuracy required.

10.3 Analysis of the samples

Follow the instrument manufacturer’s instructions for set up, conditioning, preparation and maintenance of the XRF spectrometer.

Select the required preparation method and prepare the samples. For the quantification of trace elements the pressed pellet method is recommended and for the determination of major and minor elements the fused bead method should be used.

To analyse the prepared samples, an analytical measurement method has to be defined. The measurement method describes the analytical lines to be measured and the measurement parameters

e.g. the XRF generator settings (tube voltage and current), selection of primary beam filters, targets and crystals, detector to be used, and measurement time.

The same measurement parameters used for the calibration according to 10.2 are applied to the samples.

Before analysis, quality control samples have to be measured to check the instrument stability and the quality of the calibration, in accordance to the manufacturer’s instructions.

Introduce the prepared sample into the XRF spectrometer and analyse it in accordance to the manufacturer’s instructions.

11.0 Quality control

11.1 Drift correction procedure

XRF calibrations, once established, tend to be stable over long periods of time. Small amounts of instrumental drift can be corrected by analysing stable monitor samples as frequency as performance experience indicates.

Drift correction monitors are stable beads that should contain all the elements to be determined and at concentration levels comparable to or higher than those from the samples.

The monitor samples shall be measured together with the calibration samples in order to get the initial intensities stored. When drift correction is needed, they are measured again. The initial set and the actual set of intensities are used to adjust the calibration regression. The procedure described is usually part of the instruments software.

For EDXRF spectrometers, an additional energy calibration has to be performed on a regular basis, as defined by the manufacturer’s instructions.

11.1.1 Blank test

To assess the level of impurities in the flux, carry out a blank test by preparing and analysing a blank sample using the same sample preparation procedure and the same quantities of reagents.

11.1.2 Reference materials

Verify the trueness of the results by applying the procedure to one or more reference materials not used for calibration and covering the concentration range of interest.

The element content of the reference material used shall be in accordance with the concentration range of interest.

12.0 Calculation of the result

Follow the guideline in the instrument manufacturer’s manual how to perform the regression, the background correction and the overlap correction.

The concentrations of the analytes are calculated by the software program from the measured intensities using the calibrations curves previously set-up. The results are expressed as elements in mg/kg dry matter for trace elements and as oxides in mass percentages dry matter for major and minor elements.

13.0 Test report

In addition to test results the test report shall include at least the following information:

a) the test method used, together with a reference to this document, i.e. ISO/DIS 18227:2024.

b) the description and identification of the laboratory sample;

c) which processes, procedures and apparatus were used;

d) results of the determination expressed in mg/kg dry matter or in mass percentages dry matter;

e) any details not specified in this document or which are optional, and any other factors which can have affected the results;

f) the date of receipt of laboratory sample and date(s) of performance of test;

Corrections or additions to a test report after issue shall be made only by a further document suitably marked, e.g. "Amendment/Addendum to test report serial number (or as otherwise identified)", and shall meet the relevant requirements of the preceding paragraphs.


  1. (informative)

    Semi-quantitative screening analysis of waste, sludge and soil samples
    1. Principle

The principle is identical with that described in Clause 5, however the elemental composition of the sample is determined by reference to calibration curves, previously set up by the manufacturer. This procedure is often referred to as “standardless” analysis. The method is generally applicable for the semi- quantitative determinations of elements from sodium to uranium except noble gases at levels between approximately 0,01 % and 100 %, depending on the element and the instrument used.

During the evaluation and calculation of the element concentration of the sample the various interferences e.g. spectral line overlap, matrix effects, spectral artefacts and sample preparation are all accounted for with the provided analytical program.

    1. Energy dispersive (ED) or wavelength dispersive (WD) X-ray fluorescence spectrometer

The same instruments as those described in 6.1 can be used; however a specific software package suitable to perform the XRF analysis without the use of calibration curves set up with dedicated reference samples, is applied. Most of the instruments available are delivered with pre-calibrated analytical methods. These calibrations are set up by the manufacturer with a suite of synthetic calibration samples to cover a wide concentration range on a broad spectrum of matrix types. Improvement of the accuracy can be obtained by additional analyses of sample specific reference materials and by extending the calibration for the specific needs.

NOTE Because of the differences between various models of XRF instruments, no detailed operating instructions can be provided.

The validity of the programmed calibration curves can be checked and optimized by using reference materials of a similar composition as the samples under investigation.

    1. Sample preparation

The sample preparation determines significantly the obtained quality of the XRF results. For detailed information on sample preparation procedures, refer to the flowcharts and the sample preparation techniques in Annex B.

    1. Procedure
      1. Analytical measurement conditions and calibration

All X-ray spectrometers are supplied with a spectrometer software program to operate the instrument. The software packages are manufacturer dependent and contain two major modules:

— analytical measurement program for data collection. This module controls the measurement of a sample using a certain set of measurement parameters e.g. tube setting (voltage, current), targets and crystals, detectors, and measurement times. The analytical program is always linked to a selected evaluation and calibration program. Actually, the same measurement conditions have to be applied for both the standards of the calibration curve and the samples. Because in screening analysis the measurements will be performed with the predefined analytical programs, no further detailed descriptions will be given of the analytical measurement parameters. Follow the manufacturer’s instruction for further operation and handling of the analytical software package;

— evaluation program for data processing. This module converts the measured intensities of the different element line to elemental concentrations taking all corrections into account. There are various types of evaluation programs available and each manufacturer has set up his program for data processing based on the XRF principles.

Sensitivity, instrumental detection limits and precision are instrument-dependent and should therefore be investigated and established for each individual analyte line on that particular instrument, and, if relevant, in function of matrix type and sample preparation.

      1. Validation

Prior to analysis of a sample, the available pre-calibrated analytical method has to be validated by using reference samples with a similar composition as the samples under investigation. If no reference materials with a comparable matrix are available, only a qualitative analysis with indicative concentration values can be performed.

The reference sample can consist of:

— in-house or commercially available reference materials, if possible certified, with matrices similar to that of the sample;

— synthetic samples, made by weighing the appropriate amount of each pure reagent;

— site specific or batch specific samples, similar to the matrix of the sample;

— standard addition method or spiked samples can also be used to create standards for which appropriate reference materials are not available for an element of interest. The matrix material needs to match that of the sample.

The element concentrations of these reference samples have to be known by certification or by determination with a different analytical technique.

The reference samples have to be analysed under the same analytical conditions as the sample, meaning the same sample preparation (pellet, powder etc.), the same analytical measurement method etc.

      1. Analysis

Follow the instrument’s instructions for set up, conditioning, preparation and maintenance of the XRF spectrometer.

To analyse the prepared samples, an analytical measurement method has to be defined. Depending on the type of instrument, the analytical measurement method is immediately applicable, or can be modified starting from a “master” measurement method or has to be set up using the available pre-calibrated spectral lines. The measurement method describes the analytical lines to be measured and the measurement parameters like the XRF generator settings (tube voltage and current), selection of primary beam filters, targets and crystals, detector to be used, measurement time etc.

The same measurement parameters of the analyte line used for the calibration have to be applied for the samples. The provided software program has to be capable of defining automatically for each analyte the required measurement parameters.

Before analysis, quality control samples have to be measured to check the instrument stability and the quality of the calibration, in accordance to the manufacturer’s instruction.

Introduce the prepared sample into the XRF spectrometer and analyse it with the selected analytical measurement method, in accordance to the manufacturer’s instruction.

    1. Calculation and evaluation

The software program from the measured intensity calculates the element concentrations, expressed in mg/kg dry matter for trace and minor elements and m/m % dry matter for major elements.

The effective matrix corrections have to account for low atomic mass elements like hydrogen, carbon or oxygen. As these elements are easily absorbed they have a strong influence on the matrix effects. Different approaches are applied in commercial software packages for this effect. First of all, automated correction using mass absorption coefficients can be performed. Secondly, it can be possible to enter the matrix compounds e.g. CH2 for polymers, H2O for water samples, and C for coal samples, to optimize the calculation process. All these methods allow the use of one calibration for different matrices.

NOTE In some cases, it is possible to optimize the iterative calculation process for the element concentration by inserting concentration values of specific compounds, obtained with another analytical technique.

    1. Quality control

The quality control has to be done according to Clause 11.

    1. Safety remarks

The safety remarks are the same as described in Clause 4.

    1. Test report

The test report has to be done according to Clause 13.


  1. (informative)

    Examples for operational steps of the sample preparation for soil and waste samples
    1. General

As waste samples can consist of different material types, various sample preparations have to be applied prior to the XRF screening analysis. The operational steps for different matrices are given in the flow diagrams in Figures B.1 to B.5:

— Figure B.1: Operational steps for the preparation of soil, sediment, fly ash and sludge samples;

— Figure B.2: Operational steps for the preparation of samples consisting of carbon matrices;

— Figure B.3: Operational steps for the preparation of liquid samples;

— Figure B.4: Operational steps for the preparation of paste-like materials;

— Figure B.5: Operational steps for the preparation of scrap samples.

The procedures of the different sample preparation techniques, as mentioned in Figures B.1 to B.5, are explained in the normative part (Clause 9.1 and 9.2) or otherwise given directly.

    1. Preparation of liquid samples

If the liquid sample is homogeneous, it is poured directly into a sample cup. The bottom of the sample cup is closed by a thin-film support. Close the sample cup.

    1. Preparation of powder samples

After milling/grinding of the sample to powder (6.4), the homogenized powder sample can be poured directly into a sample cup. The bottom of the sample cup is closed by a thin-film support. Press the powder slightly with a piston to form a flat and even surface and avoid any air bubbles inside the powder. Close the sample cup.

    1. Impregnation on powders
      1. Impregnation on aluminium oxide

Place 15,000 g ± 0,005 g of aluminium oxide (pre-calcined at 1 500 °C) into a HDPE bottle. Add 5,000 g ± 0,005 g of the homogeneous sample and two mixing balls. Close the bottle and shake vigorously for approximately 30 s. Empty the aluminium oxide / sample mixture from the bottle to the sample cup. Press the powder slightly with a piston to form a flat and even surface and avoid any air bubbles inside the powder.

      1. Impregnation on activated carbon

Weight 4,000 g ± 0,005 g of sample, add 2,000 g ± 0,005 g of active carbon and homogenize very well the mixture. Empty the active carbon / sample mixture from the bottle to the sample cup. The powder is slightly pressed with a piston to form a good surface to avoid any air holes on the bottom. High volatile liquids can be impregnated on active carbon.

Figure B.1 — Operational steps for the preparation of soil, sediment, fly ash and sludge samples

Figure B.2 — Operational steps for the preparation of samples consisting of carbon matrices

Figure B.3 — Operational steps for the preparation of liquid samples

Figure B.4 — Operational steps for the preparation of paste-like materials

Figure B.5 — Operational steps for the preparation of scrap samples


  1. (informative)

    Suggested analytical lines, crystals and operating conditions

Optimum excitation for element analysis can be realized by applying a suitable crystal. Table C.1 shows a list with potentially usable crystals. For all of these, the tube voltage and current need to be set carefully in order to get optimum results. The number of excitation conditions selected shall be optimized, particularly, with regard to detection sensitivity and required analysis time.

Table C.1 — Suggested analytical lines, crystals and operating conditions for wavelength dispersive XRF spectrometry

Element

Line

Crystals

kV/mA

Collimator

Detector

Na

OVO-55/PX- 1/ TlAp

30/100

coarse

FC

Mg

OVO-55/PX-1/ TlAp

30/100

coarse

FC

Al

OVO-55/PX- 1/ PET

30/100

coarse

FC

Si

OVO-55/PX-1/ PET

30/100

coarse

FC

P

Ge/PET

30/100

coarse

FC

S

Ge/PET

30/100

coarse

FC

Cl

Ge/PET

30/100

coarse

FC

K

LiF200

50/60

fine

FC

Ca

LiF200

50/60

fine

FC

Ti

LiF200

50/60

fine

FC

V

LiF200

30/100

fine

FC

Cr

LiF200

60/50

fine

FC

Mn

LiF200

60/50

fine

FC

Fe

LiF200

60/50

fine

FC

Co

LiF200

60/50

fine

SC

Ni

LiF200

60/50

fine

SC

Cu

LiF200

60/50

fine

SC

Zn

LiF200

60/50

fine

SC

As

Kβ

LiF200

60/50

60/50

fine

SC

Se

LiF200

60/50

fine

SC

Br

LiF220

60/50

fine

FC

Rb

LiF220

60/50

fine

FC

Sr

LiF200

60/50

fine

FC

Y

LiF220

60/50

fine

FC

Zr

LiF220

60/50

fine

FC

Nb

LiF220

60/50

fine

FC

Mo

LiF220

60/50

fine

FC

Ag

LiF220 PET

60/50

30/100

fine coarse

SC FC

Cd

Kα Lα

LiF220 PET

60/50

30/100

fine coarse

SC FC

Optimum excitation for trace element analysis can be realized by applying secondary or polarization targets in to the excitation beam of the spectrometer. Table C.2 shows a list with potentially usable targets. For all of these, the tube voltage and current need to be set carefully in order to get optimum results. The number of excitation conditions selected shall be optimized, particularly, with regard to detection sensitivity and required analysis time.

Table C.2 — Suggested analytical lines, targets and operating conditions for energy dispersive XRF spectrometry

Target

Target type

Elements

Mo

Secondary

Cr – Zr (K)

Hf – U (L)

Zr

Secondary

Cr – Sr (K)

Hf – Bi (L)

Al2O3

Barkla

Mo - Ba (K)

CsI

Secondary

Mo – In (K)

Pd

Secondary

Fe – Mo (K)

Hf – U (L)

B4C

Barkla

Fe – Mo (K)

Hf – U (L )

Ge

Secondary

Cr – Zn (K)

Hf – Ta (L)

Zn

Secondary

Ti – Ni (K)

Co

Secondary

K – Mn (K)

Cd – La (L)

Ti

Secondary

P – Ca (K)

Si

Secondary

Na – Al (K)

Al

Secondary

Na – Mg (K)

HOPG

Bragg

Na – Sr (K)

Y – Hf (L)

Key

(K)   Kα radiation

(L)   Lα radiation


  1. (informative)

    List of reference materials applicable for XRF analysis

Reference materials should be selected containing the elements of interest and covering the total concentration range of interest. Additionally, reference materials with a similar composition of the samples under investigation should be selected.

Table D.1 — Reference materials applicable for XRF analysis

Identification

 

Matrix

ERM-CC141

Institute for Reference Materials and Measurements

Loam soil

BCR-142R

Institute for Reference Materials and Measurements

Light sandy soil

BCR-143R

Institute for Reference Materials and Measurements

Sewage sludge soil

BCR-667

Institute for Reference Materials and Measurements

Estuarine sediment

ERM-CC690

Institute for Reference Materials and Measurements

Calcareous soil

ERM-CC144

Institute for Reference Materials and Measurements

Sewage sludge

BCR-145R

Institute for Reference Materials and Measurements

Sewage sludge

BCR-146R

Institute for Reference Materials and Measurements

Sewage sludge

BCR-038

Institute for Reference Materials and Measurements

Fly ash

GBW07401-07411

National Research Centre for Certified Reference Materials (NRCCRM), China

Soil

SARM 42

SACCRM, South Africa

Soil

GBW07301, 07302, 07304, 07305, 07306 and 07308

National Research Centre for Certified Reference Materials (NRCCRM), China

Sediments

SARM 51

SACCRM, South Africa

Sediment

SARM 52

SACCRM, South Africa

Sediment

STSD-3

Canadian Certified Reference Materials (CCRMP)

Sediment

TILL-2 and 3

Canadian Certified Reference Materials (CCRMP)

Soil

N2709a

National Institute of Standards & Technology (NIST)

Soil

N2710a

National Institute of Standards & Technology (NIST)

Soil

N2711a

National Institute of Standards & Technology (NIST)

Soil

N1633c

National Institute of Standards & Technology (NIST)

Coal fly ash

N2689

National Institute of Standards & Technology (NIST)

Coal fly ash

N2690

National Institute of Standards & Technology (NIST)

Coal fly ash

N2691

National Institute of Standards & Technology (NIST)

Coal fly ash

BAM-U110

Federal Institute for Materials Research and Testing

Contaminated Soil

NOTE This list can be incomplete.


  1. (informative)

    Validation
    1. Result of validation

An inter-laboratory comparison supported by DIN was organized by CEN/TC 292 WG 3 in May 2006/August 2006 with participants from seven member countries. For the inter-laboratory comparison, two soil and three waste samples were selected and distributed to the participants. The samples are representing a wide spectrum of contents of elements and a broad variety of the matrix composition.

— Sample 1: soil sample of a sandy-loamy arable soil taken 50 m beside a highway in Germany, homogenized and homogeneity-tested by UBA (Federal Environment Agency).

— Sample 2: sample of contaminated soil of an area subject to periodic flooding of the Saale river in Germany, sandy silt, certified as soil reference material BAM-U110[14].

— Sample 3: fly ash sample of a household waste incineration plant, homogenized and homogeneity- tested by BAM (Federal Institute for Materials Research and Testing).

— Sample 4: ink waste sample, cited by (3) as SAMPLE CEN 8/99, homogeneity-tested by EU DG JRC (European Commission’s DG Joint Research Centre) and used for validation of the digestion standards EN 13656 and EN 13657[15].

— Sample 5: electronic sludge, cited by (3) as SAMPLE CEN 9/99, homogeneity tested by EU DG JRC (European Commission’s DG Joint Research Centre) and used for validation of the digestion standards EN 13565 and EN 13657[15].

All the samples were sent to the participating laboratories as dried, fine-grinded less than 90 µm, and homogenized material.

The samples were analysed using EDXRF and WDXRF instrumentation as well. As sample preparation methods, the pressed pellet and the fused bead technique were applied depending on the available laboratory equipment. For each type of sample and each preparation method, two test samples were prepared (pellets or beads) and analysed in duplicate. Finally, for each combination four results were reported.

Fifteen volunteer laboratories received the samples and all of them transmitted data. For sample 4 and 5, no data from the fused bead preparation method were reported because the commonly used fused bead method was not applicable for this type of samples.

The evaluation of the data was done according to ISO 5725‑2; the principal steps of Figure 3 were followed strictly. According to this procedure, the data of two laboratories were removed. The maximum number of laboratories providing results for the statistical evaluation was therefore 13.

For the statistical evaluation the data of the EDXRF and the WDXRF technique were combined, resulting in up to 14 different data sets for each element. For each sample and preparation method, i.e. pressed pellet or fused bead, the data were evaluated separately. Outliers were determined and eliminated according to ISO 5725‑2 using the test of Mandel statistics combined with Grubbs test. The program used for statistical evaluation of the data and calculation of the performance characteristics was written in SPSS[16].

Tables E.1 to E.8 show the performance characteristics of the five samples. Results obtained by the pressed pellet preparation method are presented for sample 1 through 5, while for the fused bead method, only results from sample 1, 2 and 3 are available. Repeatability and reproducibility were calculated according to the definitions given by ISO 5725‑2.

Table E.1 — Validation data of sample 1 (soil) — pressed pellet method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mg/kg

mg/kg

%

mg/kg

%

mg/kg

Na

12

40

4

8 182

2 402

29 %

492

6 %

8 100

Mg

12

40

4

3 930

947

24 %

238

6 %

3 800

Al

12

44

4

45 039

3 469

8 %

784

2 %

38 000

Si

13

46

0

317 432

44 283

14 %

4 204

1 %

n.d.

P

11

44

4

1 111

409

37 %

26

2 %

820

S

13

46

0

1 011

358

35 %

47

5 %

n.d.

Cl

12

29

4

103

53

51 %

25

24 %

n.d.

K

12

46

4

18 000

2 200

12 %

240

1 %

17 000

Ca

14

48

0

5 353

1 617

30 %

149

3 %

6 400

Ti

13

46

0

1 824

368

20 %

37

2 %

1 800

V

13

50

0

38

14

38 %

4

10 %

35

Cr

12

50

2

64

20

31 %

10

16 %

50

Mn

12

46

4

362

40

11 %

6

2 %

390

Fe

12

46

4

12 190

974

8 %

135

1 %

11 000

Co

12

27

4

6

4

62 %

2

27 %

n.d.

Ni

12

47

4

15

5

31 %

1

9 %

14

Cu

12

49

4

25

7

28 %

2

7 %

34

Zn

12

48

0

99

15

15 %

3

3 %

120

As

12

44

4

8

4

47 %

1

13 %

14

Br

12

42

4

11

3

31 %

2

21 %

n.d.

Rb

12

50

4

97

9

10 %

1

1 %

97

Sr

12

50

4

79

9

11 %

1

1 %

80

Y

13

48

0

16

4

27 %

1

7 %

9

Zr

12

50

4

205

19

9 %

8

4 %

320

Nb

11

44

8

8

3

35 %

1

8 %

18

Sn

12

28

4

6

2

35 %

1

25 %

n.d.

Cs

12

26

4

10

5

47 %

3

26 %

8

Ba

12

50

4

404

64

16 %

13

3 %

360

W

12

31

4

12

15

125 %

2

17 %

8

Pb

12

50

4

78

11

14 %

3

3 %

86

Th

12

40

4

8

3

37 %

1

13 %

n.d.

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.2 — Validation data of sample 1 (soil) — fused bead method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mg/kg

mg/kg

%

mg/kg

%

mg/kg

Na

5

20

4

10 075

836

8 %

518

5 %

8 100

Mg

5

20

4

3 054

347

11 %

89

3 %

3 800

Al

5

20

0

42 233

980

2 %

775

2 %

38 000

Si

5

20

0

350 701

7 042

2 %

3 069

1 %

n.d.

P

5

20

0

707

203

29 %

58

8 %

820

K

5

20

4

19 282

464

2 %

187

1 %

17 000

Ca

6

24

4

4 952

215

4 %

67

1 %

6 400

Ti

5

20

0

1 849

109

6 %

45

2 %

1 800

V

5

10

2

35

2

6 %

4

11 %

35

Cr

5

11

0

67

15

22 %

5

7 %

50

Mn

5

16

0

362

50

14 %

14

4 %

390

Fe

5

20

0

12 013

2 058

17 %

275

2 %

11 000

Cu

5

10

0

69

14

21 %

5

7 %

34

Ba

5

12

4

344

73

21 %

25

7 %

360

Pb

5

10

2

82

1

1 %

2

2 %

86

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.3 — Validation data of sample 2 (contaminated soil) — pressed pellet method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mCERT

mg/kg

mg/kg

%

mg/kg

%

mg/kg

mg/kg

Na

13

34

4

4 564

1 478

32 %

549

12 %

4 700

n.d.

Mg

14

44

0

10 338

1 812

18 %

963

9 %

11 000

8 380

Al

12

47

8

55 589

3 792

7 %

583

1 %

67 000

50 382

Si

13

51

4

239 921

26 488

11 %

1 720

1 %

n.d.

n.d.

P

12

47

4

5 305

1 078

20 %

78

1 %

4 800

n.d.

S

13

50

4

13 665

2 181

16 %

314

2 %

n.d.

n.d.

Cl

13

51

4

1 782

395

22 %

103

6 %

n.d.

n.d.

K

12

51

8

20 451

1 648

8 %

195

1 %

19 000

20 381

Ca

14

55

4

46 924

7 751

17 %

1 065

2 %

47 000

40 638

Ti

14

51

0

3 474

521

15 %

50

1 %

3 300

3 206

V

14

52

0

65

28

42 %

12

18 %

62

68

Cr

14

55

0

257

37

15 %

8

3 %

220

230

Mn

14

51

0

681

112

17 %

10

2 %

680

621

Fe

14

51

0

30 840

3 976

13 %

404

1 %

25 000

28 229

Co

14

43

0

23

9

39 %

3

13 %

14

16

Ni

13

55

4

108

14

13 %

4

3 %

87

101

Cu

13

55

4

265

28

10 %

7

3 %

240

263

Zn

12

51

4

992

52

5 %

14

1 %

1 000

1 000

As

12

44

8

14

4

30 %

2

11 %

20

16

Se

12

19

3

2

1

27 %

1

25 %

n.d.

n.d.

Br

12

48

5

17

4

25 %

1

5 %

n.d.

n.d.

Rb

13

52

4

92

6

7 %

2

2 %

91

n.d.

Sr

13

52

4

277

22

8 %

5

2 %

280

277

Y

13

52

4

24

4

16 %

1

4 %

17

n.d.

Zr

14

52

0

306

43

14 %

8

3 %

190

n.d.

Nb

13

44

4

14

4

30 %

1

5 %

23

n.d.

Mo

13

28

4

4

1

28 %

1

20 %

n.d.

n.d.

Ag

12

26

2

5

2

33 %

1

26 %

7

n.d.

Cd

14

24

0

8

2

29 %

1

17 %

n.d.

7

Sn

13

45

1

23

7

31 %

3

13 %

51

n.d.

Sb

12

27

4

7

2

30 %

2

26 %

n.d.

n.d.

I

13

16

1

7

2

29 %

4

62 %

n.d.

n.d.

Cs

13

20

4

15

4

28 %

6

40 %

8

n.d.

Ba

14

55

0

1 545

276

18 %

44

3 %

1 400

1 488

Hg

12

47

8

52

11

21 %

2

4 %

37

52

Tl

13

16

4

3

1

33 %

n.d.

n.d.

Pb

14

55

0

197

33

17 %

5

3 %

210

197

Bi

14

19

0

2

1

52 %

1

76 %

n.d.

n.d.

Th

12

43

8

10

2

20 %

1

15 %

n.d.

n.d.

U

12

31

7

7

3

46 %

2

30 %

6

n.d.

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.4 — Validation data of sample 2 (contaminated soil) — fused bead method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mCERT

mg/kg

mg/kg

%

mg/kg

%

mg/kg

mg/kg

Na

4

19

3

5 958

582

10%

236

4 %

4 700

n.d.

Mg

5

19

0

8 658

789

9 %

340

4 %

11 000

8 380

Al

5

19

0

50 039

1 004

2 %

487

1 %

67 000

50 382

Si

5

19

0

257 753

3 623

1 %

1 435

1 %

n.d.

n.d.

P

4

19

3

3 555

177

5 %

47

1 %

4 800

n.d.

K

4

19

4

19 968

733

4 %

125

1 %

19 000

20 381

Ca

5

23

4

41 361

1 061

3 %

430

1 %

47 000

40 638

Ti

5

19

0

3 397

205

6 %

36

1 %

3 300

3 206

Cr

5

12

0

246

28

12 %

7

3 %

220

230

Mn

5

16

0

623

14

2 %

15

2 %

680

621

Fe

5

19

0

28 921

2 144

7 %

2 150

7 %

25 000

28 229

Cu

5

11

0

263

81

31 %

32

12 %

240

263

Zn

4

7

0

1 024

178

17 %

17

2 %

1 000

1 000

Ba

5

11

0

1 780

495

28 %

132

7 %

1 400

1 488

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.5 — Validation data of sample 3 (fly ash) — pressed pellet method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mg/kg

mg/kg

%

mg/kg

%

mg/kg

Na

13

43

4

14 548

3 484

24 %

515

4 %

15 000

Mg

13

44

4

13 218

2 650

20 %

1 385

10 %

12 000

Al

13

47

4

53 651

6 836

13 %

680

1 %

48 000

Si

13

51

4

176 582

25 716

15 %

1 778

1 %

n.d.

P

12

47

4

5 257

1 036

20 %

112

2 %

4 100

S

12

47

8

14 592

2 483

17 %

185

1 %

n.d.

Cl

13

51

4

5 088

831

16 %

147

3 %

n.d.

K

14

51

0

13 109

1 486

11 %

202

2 %

12 000

Ca

14

55

4

124 013

14 792

12 %

3 144

3 %

100 000

Ti

14

51

0

5 829

829

14 %

99

2 %

5 100

V

13

45

4

46

14

29 %

6

13 %

46

Cr

12

55

7

291

31

11 %

8

3 %

190

Mn

14

51

0

1 161

189

16 %

23

2 %

1 200

Fe

14

51

0

45 995

3 994

9 %

498

1 %

43 000

Co

13

42

4

28

14

49 %

6

20 %

17

Ni

13

55

4

104

19

19 %

4

4 %

100

Cu

12

55

8

1 528

114

7 %

17

1 %

1 600

Zn

11

51

8

4 342

273

6 %

81

2 %

4 400

As

13

22

4

12

8

62 %

3

27 %

16

Br

13

31

4

10

2

18 %

1

7 %

n.d.

Rb

13

52

4

51

5

10 %

2

4 %

50

Sr

12

52

8

337

23

7 %

7

2 %

350

Y

13

44

4

16

4

22 %

1

5 %

11

Zr

13

52

4

225

28

13 %

9

4 %

170

Nb

13

36

4

11

2

22 %

1

8 %

19

Mo

12

34

6

14

3

24 %

3

23 %

10

Ag

13

30

1

10

3

35 %

1

13 %

11

Cd

14

33

0

12

4

34 %

2

16 %

12

Sn

13

55

4

283

42

15 %

8

3 %

200

Sb

13

54

4

47

17

36 %

3

7 %

47

Ba

13

55

4

1 867

231

12 %

50

3 %

1 800

W

13

23

4

42

34

81 %

37

88 %

25

Pb

12

51

8

1 210

131

11 %

29

2 %

1 400

Bi

14

21

0

3

2

48 %

1

30 %

5

Th

13

34

2

14

6

44 %

3

19 %

n.d.

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.6 — Validation data of sample 3 (fly ash) — fused bead method

Element

l

n

n

mXRF

SR

VR

Sr

Vr

mICPMS

mg/kg

mg/kg

%

mg/kg

%

mg/kg

Na

5

20

4

17 940

1 402

8 %

666

4 %

15 000

Mg

5

20

0

12 848

744

6 %

400

3 %

12 000

Al

5

20

0

54 578

2 426

4 %

1 786

3 %

48 000

Si

5

20

0

197 167

5 562

3 %

3 194

2 %

n.d.

P

5

20

4

4 695

220

5 %

89

2 %

4 100

K

5

20

0

12 719

761

6 %

294

2 %

12 000

Ca

6

24

4

114 789

5 223

5 %

1 961

2 %

100 000

Ti

5

20

0

6 021

401

7 %

119

2 %

5 100

V

5

11

0

47

26

54 %

8

18 %

46

Cr

5

12

0

302

28

9 %

24

8 %

190

Mn

5

16

0

1 240

41

3 %

20

2 %

1 200

Fe

5

20

1

46 690

2 101

4 %

563

1 %

43 000

Cu

5

12

0

1 631

226

14 %

58

4 %

1 600

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXR mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.7 — Validation data of sample 4 (ink waste) — pressed pellet method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mEUJRC

mg/kg

mg/kg

%

mg/kg

%

mg/kg

mg/kg

Na

13

32

6

6 119

2 146

35 %

341

6 %

4 200

6 787

Mg

13

39

12

1 192

557

47 %

90

8 %

960

960

Al

13

41

9

2 536

683

27 %

194

8 %

3 200

2 111

Si

13

48

4

15 980

4 186

26 %

364

2 %

n.d.

n.d.

P

12

44

4

17 004

5 006

29 %

240

1 %

12 000

14 005

S

13

44

8

35 353

5 774

16 %

478

1 %

n.d.

33 032

Cl

13

48

4

4 630

1 961

42 %

1 436

31 %

n.d.

n.d.

K

13

44

8

1 032

388

38 %

44

4 %

510

965

Ca

14

52

4

124 307

57 376

46 %

5 685

5 %

110 000

107 315

Ti

13

44

0

318

112

35 %

11

3 %

340

232

V

13

32

0

23

13

55 %

5

23 %

20

15

Cr

13

52

4

5 005

1770

35 %

138

3 %

3 400

3 612

Mn

13

48

4

717

320

45 %

18

2 %

580

530

Fe

13

48

4

105 308

37 749

36 %

1 364

1 %

74 000

74 271

Co

13

38

8

30

14

49 %

5

19 %

13

14

Ni

13

37

6

25

12

50 %

3

12 %

19

23

Cu

13

52

4

16 530

6 586

40 %

163

1 %

11 000

12 429

Zn

12

48

4

1 504

584

39 %

28

2 %

1 200

1 203

Br

13

26

2

16

4

28 %

2

11 %

n.d.

n.d.

Rb

13

19

3

11

3

24 %

2

14 %

<5

n.d.

Sr

13

52

4

152

61

40 %

6

4 %

120

117

Y

13

14

2

10

2

18 %

3

31 %

<5

n.d.

Zr

13

47

4

30

13

43 %

5

16 %

41

n.d.

Mo

13

28

3

5

2

46 %

1

21 %

6

5

Sn

13

16

2

5

2

46 %

2

42 %

19

2

Ba

13

40

0

121

39

32 %

10

9 %

100

101

Pb

13

52

4

6 735

3 287

49 %

105

2 %

5 900

5 894

Th

13

20

4

26

17

67 %

3

10 %

n.d.

n.d.

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

Table E.8 —Validation data of sample 5 (electronic sludge) — pressed pellet method

Element

l

n

o

mXRF

SR

VR

Sr

Vr

mICPMS

mEUJRC

mg/kg

mg/kg

%

mg/kg

%

mg/kg

mg/kg

Na

12

36

0

16 970

7 880

46 %

1 445

9%

19 000

13 207

Mg

12

32

4

2 712

1 649

61 %

1 564

58 %

2 000

2 163

Al

12

40

0

114 509

20 552

18 %

1 926

2 %

77 000

75 505

Si

12

44

4

26 278

5 565

21 %

598

2 %

n.d.

n.d.

P

11

40

0

5 776

1 383

24 %

186

3 %

4 100

4 864

S

12

40

4

73 081

12 698

17 %

1 114

2 %

n.d.

68 041

Cl

12

44

0

7 684

2 536

33 %

130

2 %

n.d.

n.d.

K

12

44

4

3 025

790

26 %

71

2 %

2 700

3 130

Ca

12

44

4

68 820

9 948

14 %

766

1 %

75 000

54 808

Ti

12

44

4

132

41

31 %

6

5 %

110

116

V

12

17

5

7

5

67 %

11

6

Cr

12

44

0

101

28

28 %

8

8 %

110

85

Mn

12

44

4

777

245

31 %

54

7 %

610

622

Fe

12

44

4

7 016

1 150

16 %

170

2 %

6 100

5 055

Ni

12

48

0

2 133

514

24 %

59

3 %

1 700

1 730

Cu

12

48

0

106 900

29 992

28 %

1 246

1 %

78 000

95 369

Zn

11

44

8

255

102

40 %

7

3 %

230

249

Br

12

36

4

91

15

17 %

5

6 %

n.d.

n.d.

Rb

12

33

0

28

11

40 %

1

4 %

20

n.d.

Sr

12

48

4

203

50

25 %

4

2 %

210

175

Zr

12

44

4

33

17

52 %

5

16 %

34

n.d.

Nb

12

19

0

23

27

115 %

1

5 %

5

n.d.

Mo

12

35

4

9

5

59 %

2

22 %

8

4

Cd

12

17

3

21

27

125 %

7

31 %

n.d.

1

Sn

12

48

4

22 385

4 661

21 %

397

2 %

17 000

16 565

I

12

17

1

77

73

95 %

17

23 %

n.d.

n.d.

Ba

12

39

0

78

31

40 %

8

10 %

86

68

Ta

12

21

5

244

135

55 %

26

11 %

n.d.

n.d.

Pb

12

48

0

9 754

2 646

27 %

171

2 %

10 000

9 380

Th

12

26

4

34

26

77 %

3

10 %

n.d.

n.d.

Key

l number of accepted laboratories

n number of accepted results

o number of outliers removed

mXRF mean content of elements specified calculated from N data sets, in mg/kg dry matter

SR reproducibility standard deviation

Sr repeatability standard deviation

VR coefficient of variation of reproducibility

Vr coefficient of variation of repeatability

mICPMS content of elements specified determined by ICP/MS technique after HNO3/HCL/HF-digestion, in mg/kg dry matter

mcert certified content of elements according,[14] in mg/kg dry matter

mEUJRC mean content of elements specified as given by,[15] in mg/kg dry matter

n.d. not determined

NOTE Numbers given in italics indicate informative data according the certification report[2].

    1. Discussion of the results

The validation data show that in general the standard is fit for purpose. In some cases, there are rather high standard deviations (up to 50 %), especially in case of low elemental contents below 100 ppb. The fused bead sample preparation technique has been proved to be more accurate. Matrix-specific calibration increases the accuracies of the results.

Comparison of the means of the two preparation methods of sample 1 and 2 (soil) and sample 3 (fly ash) show comparable results for most elements within the calculated reproducibility. The results of sample 2 obtained by the method specified in EN 15309:2007 are in good agreement with the certified values[14].

Mean concentrations and reproducibility standard deviations of sample 4 (ink waste) are in most cases evidently higher than those reported in EN 13656 due to the results of three labs which have not been identified as outliers according to ISO 5725. The high loss on ignition (around 60 %) of this sample and the matrix properties are possible explanations.

The validation clearly indicates that for soil-like matrices the fused bead sample preparation is the method of choice for main group elements, whereas the pressed pellet method is generally applicable with good results for all elements. On the other hand, the validation proves that the fused bead sample preparation technique is not be used in case of waste-like matrices such as sludges.

Bibliography

[1] ASTM D 6052‑97, Standard Test Method for Preparation and Elemental Analysis of Liquid Hazardous Waste by Energy-Dispersive X-Ray Fluorescence

[2] DIN 51418‑1, Röntgenspektralanalyse — Röntgenemissions- und Röntgenfluoreszenz-Analyse (RFA) — Teil 1: Allgemeine Begriffe und Grundlagen

[3] DIN 51418‑2, Röntgenspektralanalyse — Röntgenemissions- und Röntgenfluorenszenz-Analyse (RFA) — Teil 2: Begriffe und Grundlagen zur Messung, Kalibrierung und Auswertung

[4] EN 13656, Characterization of waste — Microwave assisted digestion with hydrofluoic (HF), nitric (HNO3) and hydrochloric (HCI) acid mixture for subsequent determination of elements

[5] EN 13657, Characterization of waste — Digestion for subsequent determination of aqua regia soluble portion of elements

[6] IEC 111/24, Procedures for the Determination of Levels of Regulated Substances in Electrotechnical Products

[7] ISO 9516‑1:2003, Iron ores — Determination of various elements by X-ray fluorescence spectrometry — Part 1: Comprehensive procedure

[8] ISO 12980:2000, Carbonaceous materials used in the production of aluminium — Green coke and calcined coke for electrodes — Analysis using an X-ray fluorescence method

[9] ISO Guide 32, Calibration in analytical chemistry and use of certified reference materials

[10] ISO Guide 33, Use of certified reference materials

[11] ISO Guide 34, General requirements for the competence of reference material producers

[12] ISO/TC 183 N691 E, XRF Analysis of Cu, Pb, Zn, Ni ores and their related products

[13] G.R. Lachance and F. Claisse, Quantitative X-ray fluorescence analysis – Theory and application, ISBN 0‑471‑95167‑6, John Wiley & Sons, West Sussex, England, 1995.

[14] Certification Report C.R.M. BAM-U110; BERICHT zur Zertifizierung der Gesamtgehalte und der mit Königswasser extrahierbaren Gehalte der Elemente As, Cd, Co, Cr, Cu, Hg, Mn, Ni, Pb und Zn in einer Bodenprobe - Zertifiziertes Referenzmaterial BAM - U110), Bundesanstalt für Materialforschung und -prüfung (BAM) (http://www.bam.de/index_en.htm), FG I.1 - Anorganisch-chemische Analytik; Referenzmaterialien, Richard-Willstätter-Str. 11, D - 12489 Berlin, September 2006

[15] H. Muntau, M.I. Kisser, C. Yuste, A. Giove and M. Bianchi, The validation of CEN draft standards for the analysis of wastes, JRC, EUR 19680 EN, 2000.

[16] A. Bühl and P. Zöfel, SPSS 11 Einführung in die moderne Datenanalyse unter Windows, Scientific Tools, Pearson Studium, Munich, ed. 8, 2002.

[17] CEN/TR 15018, Characterization of waste — Digestion of waste samples using alkali-fusion techniques

[18] EN ISO 12677, Chemical analysis of refractory products by XRF — Fused cast bead method (ISO 12677:2003)

[19] ISO 5725‑2:1994, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method

[20] ISO 14869‑2:2002, Soil quality — Dissolution for the determination of total element content — Part 2: Dissolution by alkaline fusion

[21] EN 15002, Characterization of waste — Preparation of test portions from the laboratory sample

[22] EN 15934, Sludge, treated biowaste, soil and waste — Calculation of dry matter fraction after determination of dry residue or water content

[23] EN 16179, Sludge, treated biowaste and soil — Guidance for sample pretreatment

[24] EN ISO/IEC 17025, General requirements for the competence of testing and calibration laboratories (ISO/IEC 17025:2017)

[25] ISO 11464, Soil quality — Pretreatment of samples for physico-chemical analysis

[26] ISO 11465, Soil quality — Determination of dry matter and water content on a mass basis — Gravimetric method

[27] EN 15309:2007, Characterization of waste and soil — Determination of elemental composition by X-ray fluorescence

espa-banner